Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2642-2020Class III

LivaNova USA Sterile Disposable Connector,1/2x1/2, 20 units per case

LivaNova USA, Inc. / initiated 2019-11-18 / Terminated / root cause: packaging

LivaNova USA Inc. began a recall of LivaNova USA Sterile Disposable Connector,1/2x1/2, 20 units per case on . FDA classified it as Class III and gives the reason as "Some packages of sterile disposable connectors contained a 3/8x1/4 reducer instead of the specified 1/2x1/2 connector. No significant risk to health is expected as the issue is easily detectable and the different shape...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2642-2020
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
LivaNova USA Sterile Disposable Connector,1/2x1/2, 20 units per case
Product code
DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass
Reason for recall
Some packages of sterile disposable connectors contained a 3/8x1/4 reducer instead of the specified 1/2x1/2 connector. No significant risk to health is expected as the issue is easily detectable and the different shape would prevent use of the connector.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Affected customers were contacted by letter sent via email or by phone call on November 18th, 2019. Customers were informed that a single lot of LivaNova USA Sterile Disposable Connector item# EC2135S, lot number 1923900091, contained a 3/8 x 1/4 reducer instead of the specified 1/2 x 1/2 connector. Customers were requested to check their inventory, quarantine (and do not use) any affected devices, complete the Customer Response Form and submit the completed form by email to [email protected] to confirm quantity of affected products in the facility. Customers were informed that affected product should be returned to LivaNova, and that their LivaNova reference person would contact customers to organize the replacement of the affected products. Customers were informed that no patient injuries have been reported and that the event is unlikely to cause any harm to the patient or user. Customers were requested to ensure that this Medical Device Correction notice is communicated to all personnel within their organization who need to be aware of it, and, if a customer has transferred any of the affected devices to a third party, to please communicate this information to them and inform the LivaNova Quality Assurance Team at [email protected]. Customers with questions may contact LivaNova at (800) 986-4702 or by e-mail to [email protected]. Customers were informed that this action is being reported to the Food and Drug Administration (FDA). Adverse reactions or quality problems experienced with the use of this product may be reported to LivaNova at [email protected] or the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Quantity in commerce
140 units
Distribution
Distributed to medical facilities in 5 states including: Minnesota, Kansas, Missouri, Utah and Washington.
Product codes and lots
Item# EC2135S, Lot No. 1923900091

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2642-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2642-2020Class IIITerminatedVoluntary: Firm initiated