Computed Tomography X-ray System; Model uCT 530; Rx; UDI: (01) GTIN: 06971576831012 -
Shanghai United Imaging Healthcare Co., Ltd. began a recall of Computed Tomography X-ray System; Model uCT 530; Rx; UDI: (01) GTIN: 06971576831012 - Product Usage: intended to produce cross-sectional images of the body... on . FDA classified it as Class II and gives the reason as "The metal edge overlap of the mylar strip in the gantry was uneven, which may create a sharp edge". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2704-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Shanghai United Imaging Healthcare Co., Ltd.
- Product
- Computed Tomography X-ray System; Model uCT 530; Rx; UDI: (01) GTIN: 06971576831012 - Product Usage: intended to produce cross-sectional images of the body using computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, cardiac, and vascular).
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- The metal edge overlap of the mylar strip in the gantry was uneven, which may create a sharp edge.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- A Customer Notification Letter was sent to the affected consignees on 06/20/2020 via email. United Imaging will replace the mylar strip with an improved one on the affected products via on-site update. On 8/30/2024, a correction notice was sent identifying an additional serial number.
- Quantity in commerce
- 4
- Distribution
- US Nationwide distribution including in the states of CA, IN and TX. See uploaded distribution list for more details.
- Product codes and lots
- SN 300162, 300163, 300174, 300176
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2703-2020 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2704-2020 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2705-2020 | Class II | Ongoing | Voluntary: Firm initiated |