Z-2705-2020Class II

Computed Tomography X-ray System; Model uCT 550; Rx; UDI: (01) GTIN: 06971576831036 -

Shanghai United Imaging Healthcare Co., Ltd. / initiated 2020-06-20 / Open, Classified / root cause: design

Shanghai United Imaging Healthcare Co., Ltd. began a recall of Computed Tomography X-ray System; Model uCT 550; Rx; UDI: (01) GTIN: 06971576831036 - Product Usage: intended to produce cross-sectional images of the body... on . FDA classified it as Class II and gives the reason as "The metal edge overlap of the mylar strip in the gantry was uneven, which may create a sharp edge". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-2705-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Shanghai United Imaging Healthcare Co., Ltd.
Product
Computed Tomography X-ray System; Model uCT 550; Rx; UDI: (01) GTIN: 06971576831036 - Product Usage: intended to produce cross-sectional images of the body using computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, cardiac, and vascular).
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
The metal edge overlap of the mylar strip in the gantry was uneven, which may create a sharp edge.
Root cause tag
design
FDA root cause
Device Design
Action
A Customer Notification Letter was sent to the affected consignees on 06/20/2020 via email. United Imaging will replace the mylar strip with an improved one on the affected products via on-site update. On 8/30/2024, a correction notice was sent identifying an additional serial number.
Quantity in commerce
3
Distribution
US Nationwide distribution including in the states of CA, IN and TX. See uploaded distribution list for more details.
Product codes and lots
SN 353020, 353021, 353023

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2705-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2703-2020Class IIOngoingVoluntary: Firm initiated
Z-2704-2020Class IIOngoingVoluntary: Firm initiated
Z-2705-2020Class IIOngoingVoluntary: Firm initiated