Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2673-2020Class II

Software 1898072 IPC upgrade v 2.7.3.0 -

Medtronic Xomed, Inc. / initiated 2020-06-12 / Terminated / root cause: software

Medtronic Xomed, Inc. began a recall of Software 1898072 IPC upgrade v 2.7.3.0 - Product Usage: The IPC system is indicated for the incision/cutting removal, drilling and sawing of soft and hard... on . FDA classified it as Class II and gives the reason as "During internal testing execution of the next generation of Integrated Power Console (IPC) prototype it was noted that the M5 Microdebrider was rotating at a higher speed than the set value in the cosole". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2673-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Xomed, Inc.
Product
Software 1898072 IPC upgrade v 2.7.3.0 - Product Usage: The IPC system is indicated for the incision/cutting removal, drilling and sawing of soft and hard tissue and bone in surgical procedures.
Product code
ERL Drill, Surgical, Ent (Electric Or Pneumatic) Including Handpiece
Reason for recall
During internal testing execution of the next generation of Integrated Power Console (IPC) prototype it was noted that the M5 Microdebrider was rotating at a higher speed than the set value in the cosole.
Root cause tag
software
FDA root cause
Software design
Action
On June 12th 2020 the firm send a notification to its consignees giving then the following instructions: Actions Required: 1. Immediately locate, quarantine, and return any software cards for Software 1898072 IPC Upgrade v2.7.3.0. Instructions for return are found on the Consignee Confirmation form. 2. Check the software version currently installed on your IPC Console and complete the Customer Confirmation Form. If your IPC has software v2.7.3.0 installed, do not use this console with M5 handpiece. Please record your console serial number and the software version currently loaded. 3. Return the completed Customer Confirmation Form via email to [email protected] or regular mail to: Medtronic-Xomed 6743 Southpoint Drive N Jacksonville, FL 32216 ATTN: Mike Tomberlin FCA Coordinator 4. Contact your Medtronic Field Representative to install replacement software Version 2.7.4.0 (CFN 1898073).
Quantity in commerce
673 units
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AK, AZ, AR, CA,CO, CT, FL, GA, HI, ID,IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NY, NC, OH, OK, PA, RI, SC, TN, TX, UT, VA, WV, WA, WI. The countries of Australia, Austria, Belgium, Canada, China, Denmark, Finland, France, Germany, Greece, Italy, Japan, Luxembourg, Netherlands, New Zealand, Norway, Poland, Singapore, Spain, Sweden, Switzerland, United Kingdom.
Product codes and lots
Model: 1898072 Affected Lot Numbers: 0217355465, 0218377490, 0218687535, 0218928638, 0219310085,0219639840, 0217856852, 0218472653, 0218714701, 0218958034, 0219443322,0219639842, 0217858947,0218472654, 0218721203, 0218958035, 0219448447, 0219639844, 0217859045,0218665536, 0218745073, 0219004569,0219465587, 0218203285, 0218666231, 0218746389,0219201122,0219518520,0218263964, 0218666243, 0218864105, 0219255151, 0219550550, 0218293282, 0218687534, 0218900337,0219281168, 0219619754.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2673-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2673-2020Class IITerminatedVoluntary: Firm initiated