Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2697-2020Class II

IntelliVue MX700 patient monitor, model no. 865241

Philips North America LLC / initiated 2020-06-30 / Terminated / root cause: manufacturing process

Philips North America, LLC began a recall of IntelliVue MX700 patient monitor, model no. 865241 on . FDA classified it as Class II and gives the reason as "Some Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2697-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
IntelliVue MX700 patient monitor, model no. 865241
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
Some Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the monitor to stop operating or prevent it from being switched on.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On June 30, 2020, the firm sent Urgent Medical Device Correction letters to affected customers. Customers were instructed to always ensure that the monitor is grounded during operation with mains power. Affected products should be removed from use until the power supply of the monitor has been replaced. Philips Customer Service will contact customers when a replacement power supply is ready for installation. Philips expects all replacements to be complete within six months. If you need any further information or support concerning this issue, please contact your local Philips representative or Philips Customer Support at 1-800- 722-9377.
Quantity in commerce
28
Distribution
Domestic distribution to CA, HI, MA, NY, and UT. International distribution to Costa Rica, Ireland, Lebanon, and Poland.
Product codes and lots
Serial numbers DE58562415 DE58562344  DE58562353

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2697-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2696-2020Class IITerminatedVoluntary: Firm initiated
Z-2697-2020Class IITerminatedVoluntary: Firm initiated