Alaris System PC Unit Model 8015. modular infusion pump and monitoring system
CareFusion 303, Inc. began a recall of Alaris System PC Unit Model 8015. modular infusion pump and monitoring system on . FDA classified it as Class I and gives the reason as "If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2741-2020
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CareFusion 303, Inc
- Product
- Alaris System PC Unit Model 8015. modular infusion pump and monitoring system
- Product code
- FRN Pump, Infusion
- Reason for recall
- If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power loss may result in an interruption of patient therapy or monitoring.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 06/30/2020, Urgent Medical Device Recall Notifications were mailed to Directors of Biomedical Engineering, Nursing, Risk Management, Environmental Services. Actions for Clinical Users: This situation only occurs when the device is running on battery power. 1. If this issue occurs, immediately plug the PC unit into AC power and restart the infusions. 2. Obtain a replacement PC unit as soon as possible. 3. Ensure back-up devices are readily available when infusing critical medications. Please also ensure that you have back-up monitoring devices (e.g., EtCO2, SpO2) that are critically important to the patient's care. 4. Complete and return the Customer Response Card. Actions for Cleaning Personnel: Inspect all PC unit batteries for missing screws and washers during cleaning. Should missing screws or washers be identified, send PC unit to Biomedical Engineering for repair. Expedite the inspection of all PC unit batteries for loose or missing screws and washers. If loose or missing screws and washers are identified, please replace them and torque any loose screws or washers to 6 in/lb and perform battery conditioning. See Service Bulletin 592A. Contact recalling firm at 1-800- 482-4822 to order a package of screws and washers. Battery should be replaced every two years by qualified service personnel using authorized batteries. Actions by Recalling Firm: Battery Pack Assembly procedure was revised to add clarity for reassembly of the battery, including proper use of screws and washers. Updated Service Bulletin to be provided in August 2020. An initial quantity of replacement screws and washers will be provided for use. Future orders for replacement batteries will include screws and washers System rental providers, please provide a copy of this letter and enclosures to your customers who are renting affected devices. Recall resources/customer support: www.bd.com/alaris-system-hardware-recall Questions - Recall Support Center: 858-617-1316, GMB-Ala
- Quantity in commerce
- 824277
- Distribution
- U.S.: CO, FL, SC, NY, CA, NC, TX, IL, PA, AZ, LA, IN, MA, MO, AR, WI, HI, OR, OH, KS, MN, MI, WA, UT, MD, TN, CO, NJ, NE, ND, WV, KY, IA, MS, AL, DE, AK, VA, OK, NM, GA, CT, SD, NV, ID, MT, ME, VT, WY, NH, MP, DC, GU, PR, AS, RI. O.U.S.: TR, QA, CA, AE, KW, MX, IL, SA, CH, NL, BE, TW, ZA, PH, CN, AU, IN, FR, MY, NZ, GB, KR, PK, DE, SG, JP, CU, BH
- Product codes and lots
- All serial numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2740-2020 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-2741-2020 | Class I | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.