Alaris Pump Module Model 8100 -

CareFusion 303, Inc / initiated 2020-06-30 / Open, Classified / root cause: design

CareFusion 303, Inc. began a recall of Alaris Pump Module Model 8100 - Product Usage: is intended to provide trained healthcare caregivers a way to automate the programming of infusion parameters... on . FDA classified it as Class II and gives the reason as "LED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-2822-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
CareFusion 303, Inc
Product
Alaris Pump Module Model 8100 - Product Usage: is intended to provide trained healthcare caregivers a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps necessary to enter infusion data.
Product code
FRN Pump, Infusion
Reason for recall
LED display, which provides infusion or patient monitoring values, on the module may have segments that appear dim.
Root cause tag
design
FDA root cause
Device Design
Action
On 06/30/2020, Urgent Medical Device Recall Notifications will be mailed to Directors of Biomedical Engineering, Nursing, Risk Management, Environmental Services. Actions for Clinical User: 1. If the end user observes a dim segment on the module, replace the module when feasible. There is no impact to the infusion. 2. Complete and return the Customer Response Card. Actions by Biomedical Engineering: If the module has a dim segment, please contactrecalling firm at 1-800-482-4822 to order a replacement display board (P/N TC10012952 for LVP Display Board, P/N TC10003525 for Syringe/PCA Display Board, P/N TC10008126 for EtCO2 Display Board, P/N TC10005026 for SpO2 Display Board) at no charge or send the module to the firm's Service Depot for repair. Actions by Recalling Firm: If the module has a dim segment as described in this notification, a replacement part will be provided at no charge. On 12/16/22, Medical Device Notifications were sent to customers informing them that they may have received impacted parts for repair purposes or (b) received product with impacted parts. Customers were provided with links to the original recall notices and asked to review the information and complete the customer response form.
Quantity in commerce
1442097
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of CO, FL, SC, NY, CA, NC, TX, IL, PA, AZ, LA, IN, MA, MO, AR, WI, HI, OR, OH, NJ, TN, NE, WA, KS, MI, MD, WV, KY, UT, MN, IA, GA, AL, DE, AK, VA, OK, NM, CT, SD, NV, ID, MT, ME, VT, MS, WY, NH, MP, DC, ND, GU, PR, AS, RI and the countries of TR, QA, CA, AE, KW, MX, GI, IL, SA, CH, NL, BE, TW, ZA, PH, CN, AU, IN, FR, MY, NZ, GB, KR, PK, DE, SG, JP, CU, IT, BH.
Product codes and lots
All serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2822-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2822-2020Class IIOngoingVoluntary: Firm initiated
Z-2823-2020Class IIOngoingVoluntary: Firm initiated
Z-2824-2020Class IIOngoingVoluntary: Firm initiated
Z-2825-2020Class IIOngoingVoluntary: Firm initiated
Z-2826-2020Class IIOngoingVoluntary: Firm initiated