Z-0479-2021Class II
EMPOWR Partial Knee Peg Drill Guide; Model: 800-06-007
Encore Medical, LP / initiated 2020-06-19 / Open, Classified
Encore Medical, LP began a recall of EMPOWR Partial Knee Peg Drill Guide; Model: 800-06-007 on . FDA classified it as Class II and gives the reason as "There have been issues experienced with the peg preparation steps, specifically drilling of the Femoral posterior peg. Upon drilling, the shaft portion of the drill may bind within the Posterior Peg drill guide. The...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0479-2021
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Encore Medical, LP
- Product
- EMPOWR Partial Knee Peg Drill Guide; Model: 800-06-007.
- Product code
- HSX Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer
- Reason for recall
- There have been issues experienced with the peg preparation steps, specifically drilling of the Femoral posterior peg. Upon drilling, the shaft portion of the drill may bind within the Posterior Peg drill guide. The extended length of the drill guide and tight tolerances to ensure accurate preparation seem to be contributing to this situation. There is a potential of a delay in surgery if this issue were to recur.
- FDA root cause
- Other
- Action
- The method of notification was via e-mail to surgical sales agents, and the notices were issued on 06/18/2020. Additionally, a notification to return the affected drills for replacement were sent via e-mail to affected surgical agents on 07/20/20. The following recommendations should be communicated to the implanting surgeons: Sales Agents Action: To avoid the potential binding, the following is recommended: " Use a plunge mill approach, i.e. plunge the drill in and out of the drill guide bushing a few times slowly until the hole is fully prepared. " Use of saline or fat to keep the drill lubricated during use. " Inspect all instrumentation prior to use to identify any wear, breakage or damage. Along with the notification, the communication includes an acknowledgment form that should be completed and returned to confirm that the instructions provided have been read and understood. THE ACKNOWLEDGEMENT SHOULD BE REVIEWED AND COMPLETED BY JULY 6, 2020.For questions, contact Teffany Hutto by phone at (512) 834-6255 or by email at [email protected].
- Quantity in commerce
- 28 units
- Distribution
- US: AZ, CA, FL, GA, IL, IN, MI, MN, NY, TX, VA, and WA.. No OUS Consignees.
- Product codes and lots
- all lot numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0479-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0480-2021 | Class II | Ongoing | Voluntary: Firm initiated |