Z-0480-2021Class II

EMPOWR Partial Knee Peg Drill; Model: 800-06-008

Encore Medical, LP / initiated 2020-06-19 / Open, Classified

Encore Medical, LP began a recall of EMPOWR Partial Knee Peg Drill; Model: 800-06-008 on . FDA classified it as Class II and gives the reason as "There have been issues experienced with the peg preparation steps, specifically drilling of the Femoral posterior peg. Upon drilling, the shaft portion of the drill may bind within the Posterior Peg drill guide. The...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0480-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Encore Medical, LP
Product
EMPOWR Partial Knee Peg Drill; Model: 800-06-008.
Product code
HSX Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer
Reason for recall
There have been issues experienced with the peg preparation steps, specifically drilling of the Femoral posterior peg. Upon drilling, the shaft portion of the drill may bind within the Posterior Peg drill guide. The extended length of the drill guide and tight tolerances to ensure accurate preparation seem to be contributing to this situation. There is a potential of a delay in surgery if this issue were to recur.
FDA root cause
Other
Action
The method of notification was via e-mail to surgical sales agents, and the notices were issued on 06/18/2020. Additionally, a notification to return the affected drills for replacement were sent via e-mail to affected surgical agents on 07/20/20. The following recommendations should be communicated to the implanting surgeons: Sales Agents Action: To avoid the potential binding, the following is recommended: " Use a plunge mill approach, i.e. plunge the drill in and out of the drill guide bushing a few times slowly until the hole is fully prepared. " Use of saline or fat to keep the drill lubricated during use. " Inspect all instrumentation prior to use to identify any wear, breakage or damage. Along with the notification, the communication includes an acknowledgment form that should be completed and returned to confirm that the instructions provided have been read and understood. THE ACKNOWLEDGEMENT SHOULD BE REVIEWED AND COMPLETED BY JULY 6, 2020.For questions, contact Teffany Hutto by phone at (512) 834-6255 or by email at [email protected].
Quantity in commerce
28 units
Distribution
US: AZ, CA, FL, GA, IL, IN, MI, MN, NY, TX, VA, and WA.. No OUS Consignees.
Product codes and lots
all lot numbers