MatrixMIDFACE Screws -
Synthes Produktions GmbH began a recall of MatrixMIDFACE Screws - Product Usage: The intended use for this device is in fixation of cranial bones and facial skeleton in procedures such as craniotomies... on . FDA classified it as Class II and gives the reason as "One lot of MatrixMIDFACE Emergency Screw 1.8mm, self-tapp, L 12mm, in Clip was packaged with a body clip etched with a 3 (representing 3mm) when it should have been packaged with body clip etched with a 12 (representing...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2739-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- SYNTHES GmbH / Synthes Produktions GmbH
- Product
- MatrixMIDFACE Screws - Product Usage: The intended use for this device is in fixation of cranial bones and facial skeleton in procedures such as craniotomies, reconstructive procedures, fracture repair, osteotomies, and selective orthognathic surgery of the maxilla and chin.
- Product code
- JEY Plate, Bone
- Reason for recall
- One lot of MatrixMIDFACE Emergency Screw 1.8mm, self-tapp, L 12mm, in Clip was packaged with a body clip etched with a 3 (representing 3mm) when it should have been packaged with body clip etched with a 12 (representing 12mm).
- Root cause tag
- packaging
- FDA root cause
- Under Investigation by firm
- Action
- DePuy Synthes issued urgent field safety notice in June 26, 2020 to notify customers. The affiliates in the impacted countries will notify affected customers.
- Quantity in commerce
- 41 units
- Distribution
- International distribution including in the country of Australia, Germany, New Zealand, Slovakia, Spain, and Taiwan.
- Product codes and lots
- Lot # 41P2340
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2739-2020 | Class II | Terminated | Voluntary: Firm initiated |