Z-2736-2020Class I

Alaris Pump Module Model 8100

CareFusion 303, Inc / initiated 2020-06-30 / Terminated / root cause: design

CareFusion 303, Inc. began a recall of Alaris Pump Module Model 8100 on . FDA classified it as Class I and gives the reason as "Broken elements on pump module platen such as broken upper hinge post, lower hinge, and membrane frame can occur when the door is forcefully closed on an object, the device is dropped, or if improper cleaners are used...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2736-2020
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CareFusion 303, Inc
Product
Alaris Pump Module Model 8100
Product code
FRN Pump, Infusion
Reason for recall
Broken elements on pump module platen such as broken upper hinge post, lower hinge, and membrane frame can occur when the door is forcefully closed on an object, the device is dropped, or if improper cleaners are used on the membrane frame assembly. These may prevent devices from delivering accurate amounts of fluid, which may result in an over infusion, free-flow conditions, or under infusion.
Root cause tag
design
FDA root cause
Device Design
Action
On 06/30/2020, Urgent Medical Device Recall Notifications were mailed to Directors of Biomedical Engineering, Nursing, Risk Management, Environmental Services. Actions for Clinical Users: 1. Ensure back-up devices are readily available when infusing critical medications. 2. Follow the User Manual instructions on how to properly load the administration set into the Pump Module. 3. This issue will not cause a pump alarm so when beginning an infusion, and periodically during an infusion: a. Check the drip chambers for both primary and secondary infusions to ensure the drip rate correlates to the intended infusion rate. b. Know the expected duration of the infusion, and periodically check the amount of mediation/fluid remaining in the container. If the infusion seems to be running faster or slower than expected, immediately transfer the infusion to a different Pump Module. 4. If the pump module be dropped or severely jarred, immediately remove it from use and send it to biomedical engineering. Actions for Cleaning Personnel: During each cleaning open the door of each pump module and inspect the platen, hinges, and membrane frame for cracks or other damage. If damage is found, do not use the device and remove it from service and send to biomedical engineering. Actions for Biomedical Engineering: Expedite inspection. Verify the membrane frame assembly is not loose, and that the seal is not torn. Replace any platen or membrane frame assembly that is broken, cracked or damaged. If damage is found, do not use. Actions by Recalling Firm: Revised the Technical Service Manual and is offering a Medication Safety program. System rental providers, please provide a copy of this letter and enclosures to your customers who are renting affected devices. Recall resources/customer support: www.bd.com/alaris-system-hardware-recall Questions - Recall Support Center: 858-617-1316, [email protected]
Quantity in commerce
1093588
Distribution
US: CA, FL, IN, TX, NC, LA, SC, VA, PA, OK, AZ, IL, KY, UT, AL, GA, TN, MA, MI, OH, AK, OR, MS, CO, NY, WI, KS, MO, MD, NV, AR, AS, ID, VT, NE, NJ, DC, HI, CT, MT, WV, NH, WA, DE, ME, NM, MN, WY, IA, SD, ND, MP, PR, GU. OUS: NL, MX, AU, MY, NZ, ZA, GB, FR, CH, CN, SG, TW, CA, IL, KW, QA, DE, BH, IN, SA, GI, AE, PK, PH, BE
Product codes and lots
All serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2736-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2736-2020Class ITerminatedVoluntary: Firm initiated