DCA Vantage Handheld Barcode Scanner - Zebra Model - Model # DS4308 - HC0015BZZWW external accessory to DCA...
Siemens Healthcare Diagnostics, Inc. began a recall of DCA Vantage Handheld Barcode Scanner - Zebra Model - Model # DS4308 - HC0015BZZWW external accessory to DCA Vantage Analyzer US on . FDA classified it as Class II and gives the reason as "If DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integrity of read barcode data using the check digit. Also, if DCA Vantage Analyzer is configured to run Code 39...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2747-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- DCA Vantage Handheld Barcode Scanner - Zebra Model - Model # DS4308 - HC0015BZZWW external accessory to DCA Vantage Analyzer US
- Product code
- LCP Assay, Glycosylated Hemoglobin
- Reason for recall
- If DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integrity of read barcode data using the check digit. Also, if DCA Vantage Analyzer is configured to run Code 39 with check digit and an operator erroneously scans a Code 39 barcode without a check digit, the scanner incorrectly sends the code and the last character is removed.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- Siemens Healthcare Diagnostics issues a field safety notice date June 2020 to customers
- Quantity in commerce
- 1800 units
- Distribution
- Nationwide
- Product codes and lots
- DCA Vantage Handheld Barcode Scanner #11306432 Model: DS4308 - HC0015BZZWW
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2747-2020 | Class II | Terminated | Voluntary: Firm initiated |