Z-2688-2020Class II

syngo.via RT Image Suite with software versions syngo.via VB30 or VB40

Siemens Medical Solutions USA, Inc. / initiated 2020-06-24 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of syngo.via RT Image Suite with software versions syngo.via VB30 or VB40 on . FDA classified it as Class II and gives the reason as "If the user modifies for any reason (e.g. reduction of artifacts) the original image orientation of a standard MR protocol to acquire images in a different orientation for further processing in Synthetic CT , the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2688-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
syngo.via RT Image Suite with software versions syngo.via VB30 or VB40
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
If the user modifies for any reason (e.g. reduction of artifacts) the original image orientation of a standard MR protocol to acquire images in a different orientation for further processing in Synthetic CT , the software does not recognize the adapted acquisition plane. This may result in images with wrong geometry. When this distortion remains unnoticed and the images are subsequently exported to a treatment planning system (TPS), an incorrect calculated radiation treatment plan cannot be excluded. The occurence of this issue is very unlikely and has never been reported so far.
Root cause tag
software
FDA root cause
Software design
Action
Siemens Healthcare Diagnostics sent a letter notification to customers dated June 23, 2020. The described issue will be solved with a service pack for software versions syngo.via VB30 and syngo.via VB40. Depending on your current software version, the service pack VB30A_HF08 will be installed with update SY039/20/S or the service pack VB40B_HF01 with update SY040/20/S as soon as it is available.
Quantity in commerce
21 units
Distribution
US Nationwide distribution including in the states of AZ, MA, WI, FL, PA. TX, MD, LA, TX, MN, MI, NY, SC, OH, GA, CA, IL, MO, CA, TX.
Product codes and lots
syngo.via VB30 or VB40: All Lots Model #: 10496180

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2688-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2688-2020Class IITerminatedVoluntary: Firm initiated