syngo.via RT Image Suite with software versions syngo.via VB30 or VB40
Siemens Medical Solutions USA, Inc began a recall of syngo.via RT Image Suite with software versions syngo.via VB30 or VB40 on . FDA classified it as Class II and gives the reason as "If the user modifies for any reason (e.g. reduction of artifacts) the original image orientation of a standard MR protocol to acquire images in a different orientation for further processing in Synthetic CT , the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2688-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- syngo.via RT Image Suite with software versions syngo.via VB30 or VB40
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- If the user modifies for any reason (e.g. reduction of artifacts) the original image orientation of a standard MR protocol to acquire images in a different orientation for further processing in Synthetic CT , the software does not recognize the adapted acquisition plane. This may result in images with wrong geometry. When this distortion remains unnoticed and the images are subsequently exported to a treatment planning system (TPS), an incorrect calculated radiation treatment plan cannot be excluded. The occurence of this issue is very unlikely and has never been reported so far.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens Healthcare Diagnostics sent a letter notification to customers dated June 23, 2020. The described issue will be solved with a service pack for software versions syngo.via VB30 and syngo.via VB40. Depending on your current software version, the service pack VB30A_HF08 will be installed with update SY039/20/S or the service pack VB40B_HF01 with update SY040/20/S as soon as it is available.
- Quantity in commerce
- 21 units
- Distribution
- US Nationwide distribution including in the states of AZ, MA, WI, FL, PA. TX, MD, LA, TX, MN, MI, NY, SC, OH, GA, CA, IL, MO, CA, TX.
- Product codes and lots
- syngo.via VB30 or VB40: All Lots Model #: 10496180
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2688-2020 | Class II | Terminated | Voluntary: Firm initiated |