Beacon EUS Access System Preloaded Access 135degree Needle sterile, single-use, endoscopic ultrasound device...
Covidien Llc began a recall of Beacon EUS Access System Preloaded Access 135degree Needle sterile, single-use, endoscopic ultrasound device consisting of a sharp stylet and an 18.5-gauge... on . FDA classified it as Class II and gives the reason as "This voluntary recall is being conducted following three customer reports of the device cannula detaching during a procedure. In one case, the detached cannula was not retrieved during the procedure. Potential risks...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2769-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Beacon EUS Access System Preloaded Access 135degree Needle sterile, single-use, endoscopic ultrasound device consisting of a sharp stylet and an 18.5-gauge, echo-enhanced, tip-shaped, nitinol cannula intended to pierce tissue and then act as an access channel to facilitate guidewire placement to locations of the gastrointestinal tract.
- Product code
- FCG Biopsy Needle
- Reason for recall
- This voluntary recall is being conducted following three customer reports of the device cannula detaching during a procedure. In one case, the detached cannula was not retrieved during the procedure. Potential risks associated with a detached cannula include delay in treatment, biliary leak, pain and retained device fragments.
- FDA root cause
- Under Investigation by firm
- Action
- Medtronic sent a letter notification to customers dated July 06, 2020 recalling effected lots of product. This letter instructed customers to discontinue use of the product and quarantine effected inventory. Customers were also instructed to return effected product and complete recall response forms accordingly.
- Quantity in commerce
- 269 units
- Distribution
- world wide distribution
- Product codes and lots
- Model # DSA-135-01 UDI: 10884521663527 Lot #: F2510371X, F2510861X, F2512341X, F2512383X, F2512493X, F2513453X
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2768-2020 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2769-2020 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY