Calcaneus Plate, Extensile, Small, Left, Catalog Number CPF 101 -
Flower Orthopedics Corporation began a recall of Calcaneus Plate, Extensile, Small, Left, Catalog Number CPF 101 - Product Usage: intended to be used for internal fixation of fractures and reconstruction of... on . FDA classified it as Class II and gives the reason as "The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium)". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.
Recall record
- Recall number
- Z-2782-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FLOWER ORTHOPEDICS Corporation dba Conventus Flower
- Product
- Calcaneus Plate, Extensile, Small, Left, Catalog Number CPF 101 - Product Usage: intended to be used for internal fixation of fractures and reconstruction of bones of the rearfoot, including the calcaneus.
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium).
- Root cause tag
- supplier
- FDA root cause
- Mixed-up of materials/components
- Action
- The Regional Business Director notified the customer by telephone and informed them that there was an issue with the newly released calcaneus plates. The 6 units were requested to be returned to Flower Orthopedics.
- Quantity in commerce
- 25
- Distribution
- US Nationwide distribution including in the states of AZ, FL, and OH.
- Product codes and lots
- Lot Number 1909660028; UDI 00840118115643
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2782-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2783-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2784-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2785-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2786-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2787-2020 | Class II | Terminated | Voluntary: Firm initiated |