Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2785-2020Class II

Calcaneus Plate, Extensile, Small, Right, Catalog Number CPF 201 -

FLOWER ORTHOPEDICS Corporation dba Conventus Flower / initiated 2020-07-08 / Terminated / root cause: supplier

Flower Orthopedics Corporation began a recall of Calcaneus Plate, Extensile, Small, Right, Catalog Number CPF 201 - Product Usage: intended to be used for internal fixation of fractures and reconstruction of... on . FDA classified it as Class II and gives the reason as "The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium)". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-2785-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FLOWER ORTHOPEDICS Corporation dba Conventus Flower
Product
Calcaneus Plate, Extensile, Small, Right, Catalog Number CPF 201 - Product Usage: intended to be used for internal fixation of fractures and reconstruction of bones of the rearfoot, including the calcaneus.
Product code
HRS Plate, Fixation, Bone
Reason for recall
The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium).
Root cause tag
supplier
FDA root cause
Mixed-up of materials/components
Action
The Regional Business Director notified the customer by telephone and informed them that there was an issue with the newly released calcaneus plates. The 6 units were requested to be returned to Flower Orthopedics.
Quantity in commerce
25
Distribution
US Nationwide distribution including in the states of AZ, FL, and OH.
Product codes and lots
Lot Number 1909660034; UDI 00840118115674

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2785-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2782-2020Class IITerminatedVoluntary: Firm initiated
Z-2783-2020Class IITerminatedVoluntary: Firm initiated
Z-2784-2020Class IITerminatedVoluntary: Firm initiated
Z-2785-2020Class IITerminatedVoluntary: Firm initiated
Z-2786-2020Class IITerminatedVoluntary: Firm initiated
Z-2787-2020Class IITerminatedVoluntary: Firm initiated