Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2810-2020Class II

Phillips 3.7V Rechargeable Li-Polymer Battery, Product #: 989803191341 -

Philips North America LLC / initiated 2020-06-18 / Terminated / root cause: manufacturing process

Philips North America, LLC began a recall of Phillips 3.7V Rechargeable Li-Polymer Battery, Product #: 989803191341 - Product Usage: This battery is used in the Expression MR400 MRI Patient Monitoring... on . FDA classified it as Class II and gives the reason as "A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2810-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Phillips 3.7V Rechargeable Li-Polymer Battery, Product #: 989803191341 - Product Usage: This battery is used in the Expression MR400 MRI Patient Monitoring System (Model MR400) is intended to monitor vital signs for patients undergoing MRI procedures and to provide signals for synchronization for the MRI scanner.
Product code
MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)
Reason for recall
A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights solid red and the battery will not charge. This may result from poor electrical contact between the battery and charger.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Philips Healthcare sent a customer notification letter dated June 2020 explaining the potential issues observed with their products. This letter instructs the consumer to contact the company with any reports of issues associated with this failure.
Quantity in commerce
24,545 Units
Distribution
World wide distribution. Firm reports distributing more product then quantity manufactured. Clarification requested. And received. Firm made error in counting number of batteries distributed.
Product codes and lots
Product #: 989803191341 Manufacturing date codes 2018-12-DD to 2020-02-DD

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2810-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2810-2020Class IITerminatedVoluntary: Firm initiated
Z-2811-2020Class IITerminatedVoluntary: Firm initiated
Z-2812-2020Class IITerminatedVoluntary: Firm initiated