Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2752-2020Class II

Randox Lipase Assay: Lipase Colorimetric R1 Buffer -

Randox Laboratories, Limited / initiated 2020-06-26 / Terminated

RANDOX LABORATORIES, LTD. began a recall of Randox Lipase Assay: Lipase Colorimetric R1 Buffer - Product Usage: A Lipase test system is a device intended for the quantitative in vitro determination of... on . FDA classified it as Class II and gives the reason as "Randox have confirmed imprecision of quality control and patient samples when using specific Randox Lipase buffer batches". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2752-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Randox Laboratories, Limited
Product
Randox Lipase Assay: Lipase Colorimetric R1 Buffer - Product Usage: A Lipase test system is a device intended for the quantitative in vitro determination of Lipase in human serum and plasma.
Product code
CHI Lipase-Esterase, Enzymatic, Photometric, Lipase
Reason for recall
Randox have confirmed imprecision of quality control and patient samples when using specific Randox Lipase buffer batches.
FDA root cause
Unknown/Undetermined by firm
Action
Randox sent an urgent medical device correction notification dated June 22, 2020 explaining the current failure with select lots of Lipase Colorimetric products. The firm advised customers to scrap all effected product, keep proof of destruction, review replacements needs, review results of patients generated using affected batches and return the firm's response card.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of Puerto Rico and WV.
Product codes and lots
Catalogue Numbers: LI011/004/UL, LI011/000/UL Lot #: 1048LI, 1050LI

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2752-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2751-2020Class IITerminatedVoluntary: Firm initiated
Z-2752-2020Class IITerminatedVoluntary: Firm initiated
Z-2753-2020Class IITerminatedVoluntary: Firm initiated