NextStep Retrograde Replacement Hub Set -
Arrow International Inc began a recall of NextStep Retrograde Replacement Hub Set - Product Usage: is indicated for use in the replacement of a NextStep Retrograde hub connection assembly that has been... on . FDA classified it as Class II and gives the reason as "The lidstock states the incorrect expiration date for the product. This issue could result in use of a device that is expired which could potentially lead to an increased risk of infection or other complications...". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-2804-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arrow International LLC (subsidiary of Teleflex Incorporated)
- Product
- NextStep Retrograde Replacement Hub Set - Product Usage: is indicated for use in the replacement of a NextStep Retrograde hub connection assembly that has been damaged.
- Product code
- NFK Kit, Repair, Catheter, Hemodialysis
- Reason for recall
- The lidstock states the incorrect expiration date for the product. This issue could result in use of a device that is expired which could potentially lead to an increased risk of infection or other complications. Sterility, biocompatibility, safety, or efficacy of the kits and their components are not assured beyond the correct expiration date.
- Root cause tag
- sterility
- FDA root cause
- Incorrect or no expiration date
- Action
- Arrow sent a customer notification letter dated July 6, 2020. This letter informed customers of the affected product, asked them to check and remove inventory, and complete the response form.
- Distribution
- World wide distribution
- Product codes and lots
- Product Code: NRH-15192 Lot #: 13F19H0454
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2801-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2802-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2803-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2804-2020 | Class II | Terminated | Voluntary: Firm initiated |