Z-2792-2020Class II

BRAINLAB Curve; Model/catalogue (article) numbers for Curve: 19901; 19901B; 19900; 19905 UDI (GTIN) for...

Brainlab, Inc. / initiated 2020-07-16 / Terminated / root cause: manufacturing process

Brainlab AG began a recall of BRAINLAB Curve; Model/catalogue (article) numbers for Curve: 19901; 19901B; 19900; 19905 UDI (GTIN) for Curve: 04056481138011 on . FDA classified it as Class II and gives the reason as "Incorrect manufacturer calibration". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2792-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Brainlab, Inc.
Product
BRAINLAB Curve; Model/catalogue (article) numbers for Curve: 19901; 19901B; 19900; 19905 UDI (GTIN) for Curve: 04056481138011
Product code
OLO Orthopedic Stereotaxic Instrument
Reason for recall
Incorrect manufacturer calibration
Root cause tag
manufacturing process
FDA root cause
Process control
Action
1. Refer to the appendix for the list of Brainlab Curve or Kick Navigation Systems with an affected camera installed to determine if your navigation system is affected. 2. For Cranial or ENT navigation software in combination with an affected system camera installed: Do not use the Z-touch Laser Pointer for patient registration. Use other instruments with the standard instrument pivoting method for Surface Matching or Landmark Registration, or Automatic Image Registration (AIR) if available. 3. Please continue to follow the instructions and warnings as described in the user guide. Especially relevant is the following warning, in addition to the instructions on the registration verification page appearing in the Cranial or ENT software: Verify accuracy at multiple anatomical landmarks, especially in the region of interest, as it may differ from the accuracy verified on the skin surface. If the region of interest is not accessible, verify in areas as close as possible to the region of interest. 4. In order to avoid nuisance with Spine & Trauma 3D or Orthopedic Navigation software in combination with an affected system camera installed: Do not attempt to use the optional Disposable Clip-On Remote Control. For patient registration, acquire registration points with the standard instrument pivoting method; or for Spine & Trauma 3D use Automatic Image Registration (AIR) if available.
Quantity in commerce
18 units
Distribution
USA: CA, TX, RI OUS: Armenia; Australia; Brazil; Bulgaria; Canada; Finland; France; Germany; Hong Kong; India; Israel; Italy; Japan; Mongolia; Morocco; Spain
Product codes and lots
CURVE 1.1 DUAL NAVIGATION STATION, Serial Numbers:  3911619001-19901, S/N of affected NDI Camera: P7-21527  1173019001-19901B, S/N of affected NDI Camera: P7-21173  1444519001-19901B, S/N of affected NDI Camera: P7-21183  1937319001-19901B, S/N of affected NDI Camera: P7-21325  2201419001-19901B, S/N of affected NDI Camera: P7-21324  2456219001-19901B, S/N of affected NDI Camera: P7-21382  3524219001-19901B, S/N of affected NDI Camera: P7-21414  5379420001-19901B, S/N of affected NDI Camera: P7-21830  5609520001-19901B, S/N of affected NDI Camera: P7-21892  6968320001-19901B, S/N of affected NDI Camera: P7-20390  4593720001-19901B, S/N of affected NDI Camera: P7-21655  7850820001-19901B, S/N of affected NDI Camera : P7-22083  6393920001-19901B, S/N of affected NDI Camera: P7-21895  3844314001-19901B, S/N of affected NDI Camera: P7-21449   CURVE DUAL DISPLAY NAVIGATION STATION, Serial Numbers:  0824912001-19900, S/N of affected NDI Camera: P7-21890  3894613001-19900, S/N of affected NDI Camera: P7-21780  4387415001-19900, S/N of affected NDI Camera:P7-07290  CURVE SINGLE DISPLAY NAVIGATION STATION, Serial Number: 5762713001-19905, S/N of affected NDI Camera: P7-10719

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2792-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2792-2020Class IITerminatedVoluntary: Firm initiated
Z-2793-2020Class IITerminatedVoluntary: Firm initiated