Z-2793-2020Class II

BRAINLAB Kick 2; Model/catalogue (article) numbers for Kick: 18170; 18070 UDI (GTIN) for Kick: 04056481006655

Brainlab, Inc. / initiated 2020-07-16 / Terminated / root cause: manufacturing process

Brainlab AG began a recall of BRAINLAB Kick 2; Model/catalogue (article) numbers for Kick: 18170; 18070 UDI (GTIN) for Kick: 04056481006655 on . FDA classified it as Class II and gives the reason as "Incorrect manufacturer calibration". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2793-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Brainlab, Inc.
Product
BRAINLAB Kick 2; Model/catalogue (article) numbers for Kick: 18170; 18070 UDI (GTIN) for Kick: 04056481006655
Product code
OLO Orthopedic Stereotaxic Instrument
Reason for recall
Incorrect manufacturer calibration
Root cause tag
manufacturing process
FDA root cause
Process control
Action
1. Refer to the appendix for the list of Brainlab Curve or Kick Navigation Systems with an affected camera installed to determine if your navigation system is affected. 2. For Cranial or ENT navigation software in combination with an affected system camera installed: Do not use the Z-touch Laser Pointer for patient registration. Use other instruments with the standard instrument pivoting method for Surface Matching or Landmark Registration, or Automatic Image Registration (AIR) if available. 3. Please continue to follow the instructions and warnings as described in the user guide. Especially relevant is the following warning, in addition to the instructions on the registration verification page appearing in the Cranial or ENT software: Verify accuracy at multiple anatomical landmarks, especially in the region of interest, as it may differ from the accuracy verified on the skin surface. If the region of interest is not accessible, verify in areas as close as possible to the region of interest. 4. In order to avoid nuisance with Spine & Trauma 3D or Orthopedic Navigation software in combination with an affected system camera installed: Do not attempt to use the optional Disposable Clip-On Remote Control. For patient registration, acquire registration points with the standard instrument pivoting method; or for Spine & Trauma 3D use Automatic Image Registration (AIR) if available.
Quantity in commerce
9 unit
Distribution
USA: CA, TX, RI OUS: Armenia; Australia; Brazil; Bulgaria; Canada; Finland; France; Germany; Hong Kong; India; Israel; Italy; Japan; Mongolia; Morocco; Spain
Product codes and lots
KICK 2 NAVIGATION STATION, Serial Numbers:   0164217001-18170, S/N of affected NDI Camera: P7-12520  2703119001-18170, S/N of affected NDI Camera:P7-20938  2933119001-18170, S/N of affected NDI Camera:P7-21378  6171619001-18170, S/N of affected NDI Camera:P7-18830  8745417001-18170, S/N of affected NDI Camera:P7-15162  9178819001-18170, S/N of affected NDI Camera:P7-16025  9607118001-18170, S/N of affected NDI Camera:P7-13227  5469918001-18170, S/N of affected NDI Camera: P7-14965  KICK NAVIGATION STATION, Serial Number: 1190017001-18070, S/N of affected NDI Camera: P7-06110

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2793-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2792-2020Class IITerminatedVoluntary: Firm initiated
Z-2793-2020Class IITerminatedVoluntary: Firm initiated