Z-2793-2020Class II
BRAINLAB Kick 2; Model/catalogue (article) numbers for Kick: 18170; 18070 UDI (GTIN) for Kick: 04056481006655
Brainlab, Inc. / initiated 2020-07-16 / Terminated / root cause: manufacturing process
Brainlab AG began a recall of BRAINLAB Kick 2; Model/catalogue (article) numbers for Kick: 18170; 18070 UDI (GTIN) for Kick: 04056481006655 on . FDA classified it as Class II and gives the reason as "Incorrect manufacturer calibration". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2793-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Brainlab, Inc.
- Product
- BRAINLAB Kick 2; Model/catalogue (article) numbers for Kick: 18170; 18070 UDI (GTIN) for Kick: 04056481006655
- Product code
- OLO Orthopedic Stereotaxic Instrument
- Reason for recall
- Incorrect manufacturer calibration
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- 1. Refer to the appendix for the list of Brainlab Curve or Kick Navigation Systems with an affected camera installed to determine if your navigation system is affected. 2. For Cranial or ENT navigation software in combination with an affected system camera installed: Do not use the Z-touch Laser Pointer for patient registration. Use other instruments with the standard instrument pivoting method for Surface Matching or Landmark Registration, or Automatic Image Registration (AIR) if available. 3. Please continue to follow the instructions and warnings as described in the user guide. Especially relevant is the following warning, in addition to the instructions on the registration verification page appearing in the Cranial or ENT software: Verify accuracy at multiple anatomical landmarks, especially in the region of interest, as it may differ from the accuracy verified on the skin surface. If the region of interest is not accessible, verify in areas as close as possible to the region of interest. 4. In order to avoid nuisance with Spine & Trauma 3D or Orthopedic Navigation software in combination with an affected system camera installed: Do not attempt to use the optional Disposable Clip-On Remote Control. For patient registration, acquire registration points with the standard instrument pivoting method; or for Spine & Trauma 3D use Automatic Image Registration (AIR) if available.
- Quantity in commerce
- 9 unit
- Distribution
- USA: CA, TX, RI OUS: Armenia; Australia; Brazil; Bulgaria; Canada; Finland; France; Germany; Hong Kong; India; Israel; Italy; Japan; Mongolia; Morocco; Spain
- Product codes and lots
- KICK 2 NAVIGATION STATION, Serial Numbers: 0164217001-18170, S/N of affected NDI Camera: P7-12520 2703119001-18170, S/N of affected NDI Camera:P7-20938 2933119001-18170, S/N of affected NDI Camera:P7-21378 6171619001-18170, S/N of affected NDI Camera:P7-18830 8745417001-18170, S/N of affected NDI Camera:P7-15162 9178819001-18170, S/N of affected NDI Camera:P7-16025 9607118001-18170, S/N of affected NDI Camera:P7-13227 5469918001-18170, S/N of affected NDI Camera: P7-14965 KICK NAVIGATION STATION, Serial Number: 1190017001-18070, S/N of affected NDI Camera: P7-06110
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2792-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2793-2020 | Class II | Terminated | Voluntary: Firm initiated |