Z-2879-2020Class II
VITROS XT7600 Integrated System, Product code 6844461
Ortho-Clinical Diagnostics, Inc. / initiated 2020-07-20 / Terminated / root cause: manufacturing process
Ortho Clinical Diagnostics began a recall of VITROS XT7600 Integrated System, Product code 6844461 on . FDA classified it as Class II and gives the reason as "When processing MicroSlides, slides that utilize Slot 18 in the CM rotor may be mis-positioned. As a result, certain VITROS MicroSlides may produce erroneous results". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2879-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS XT7600 Integrated System, Product code 6844461
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- When processing MicroSlides, slides that utilize Slot 18 in the CM rotor may be mis-positioned. As a result, certain VITROS MicroSlides may produce erroneous results.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm notified customers through Urgent Product Correction Notification letters on July 20, 2020. Customers were informed of the potential issue with incubator slot 18 of affected devices. Customers were provided with a table of the eight assays with biases which do not align with the assay's intended clinical use. Customers who had already been identified as experiencing this issue through their e-connected systems, or who did not have e-connected systems, received a letter stating that the recalling firm would be replacing the CM Rotor to resolve the issue. Customers who had e-connected systems but who had not yet been identified as experiencing this issue were informed that Ortho would continue to monitor their system for an increase in the frequency of condition codes related to slide misalignment. If impacted, those customers would also receive a replacement CM Rotor. Customer actions: " If you suspect an erroneous result has occurred or you observe an increase in the frequency of TH4-60J/TH4-63J condition codes, contact Ortho Care Technical Solutions Center. " Review the enclosed Assay Observed Bias sheet for VITROS Slides assays used by your laboratory. Discuss any concerns with your Laboratory Medical Director. " Complete the Confirmation of Receipt form and return to the firm " Post this notification by each VITROS XT 3400 and/or XT 7600 System in your facility or with the user documentation.
- Quantity in commerce
- pending
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
- Product codes and lots
- Serial numbers J76000050 - 76000566
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2878-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2879-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.