Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2927-2020Class II

Dimension Vista High Sensitivity Troponin I (TNIH) Flex reagent cartridge-IVD-In vitro diagnostic use in the...

Siemens Healthcare Diagnostics Inc. / initiated 2020-07-24 / Terminated

Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista High Sensitivity Troponin I (TNIH) Flex reagent cartridge-IVD-In vitro diagnostic use in the quantitative measurement of cardiac troponin I in... on . FDA classified it as Class II and gives the reason as "Dimension Vista High Sensitivity Troponin I (TNIH) may exhibit a negative bias with patient samples". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2927-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Dimension Vista High Sensitivity Troponin I (TNIH) Flex reagent cartridge-IVD-In vitro diagnostic use in the quantitative measurement of cardiac troponin I in human plasma Catalog Number K6427, SMN# 104471067
Product code
MMI Immunoassay Method, Troponin Subunit
Reason for recall
Dimension Vista High Sensitivity Troponin I (TNIH) may exhibit a negative bias with patient samples
FDA root cause
Under Investigation by firm
Action
Siemens Healthcare issued Urgent Medical Device Correction (UMDC # VC-20-03.A.US) and Urgent Field Safety Notice (UFSN # VC-20-03.A.OUS) on July 24, 2020 to Dimension Vista customers who received the Dimension Vista TNIH reagent to notify them of the issue, the potential risk to health, and actions to be taken by the customer. Customers were instructed to: "Review the letter with their Medical Director Use lot specific correlation factors with lots 20008BB, 20035BC, and 20135BB. Siemens is working to restore assay performance, until such time, subsequent lots may contain an alert card in the carton containing lot specific correlation factors ."Complete and return the Field Correction Effectiveness Check Form attached to the letter within 30 days. "If they have received any complaints of illness or adverse events associated with the product listed in Table 1 of the letter, immediately contact their local Siemens Remote Services Center or their local Siemens Technical Support Representative. Questions, please contact your Siemens Remote Services Center or your local Siemens Technical Support.
Quantity in commerce
US: 1239 units OUS: 3387
Distribution
Nationwide
Product codes and lots
Lot # 20008BB UDI: 0084276803771920135BB21031010471067840; Lot #20035BC UDI:0084276803771920035BC20113010471067840; Lot #20135BB UDI: 0084276803771920008BB20110310471067840

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2927-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2927-2020Class IITerminatedVoluntary: Firm initiated