Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2871-2020Class II

ALTERA Spacer ALTERA Spacer, 10x36, 12-16mm, 15

Globus Medical, Inc. / initiated 2020-07-22 / Terminated / root cause: manufacturing process

Globus Medical, Inc. began a recall of ALTERA Spacer ALTERA Spacer, 10x36, 12-16mm, 15 on . FDA classified it as Class II and gives the reason as "Specific lots of ALTERA implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2871-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Globus Medical, Inc.
Product
ALTERA Spacer ALTERA Spacer, 10x36, 12-16mm, 15
Product code
MAX Intervertebral Fusion Device With Bone Graft, Lumbar
Reason for recall
Specific lots of ALTERA implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
A customer letter dated July 22, 2020 was mailed to customers. In this letter customers are instructed to immediately examine their inventory and determine if they have any of the affected implants. Alternatively, they may return all ALTERA inventory and Globus will sort out the affected implants and replace inventory with conforming implants. Report any adverse effects believed to be associated with the product to Globus Medical immediately by filling out the attached Response Card.
Distribution
Worldwide Distribution
Product codes and lots
Part #: 1124.1235; Lot # / UDI #: BAX103WE / (01)00889095069976(10)BAX103WE; BAX227VE / (01)00889095069976(10)BAX227VE

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2871-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2850-2020Class IITerminatedVoluntary: Firm initiated
Z-2851-2020Class IITerminatedVoluntary: Firm initiated
Z-2852-2020Class IITerminatedVoluntary: Firm initiated
Z-2853-2020Class IITerminatedVoluntary: Firm initiated
Z-2854-2020Class IITerminatedVoluntary: Firm initiated
Z-2855-2020Class IITerminatedVoluntary: Firm initiated
Z-2856-2020Class IITerminatedVoluntary: Firm initiated
Z-2857-2020Class IITerminatedVoluntary: Firm initiated
Z-2858-2020Class IITerminatedVoluntary: Firm initiated
Z-2859-2020Class IITerminatedVoluntary: Firm initiated
Z-2860-2020Class IITerminatedVoluntary: Firm initiated
Z-2861-2020Class IITerminatedVoluntary: Firm initiated
Z-2862-2020Class IITerminatedVoluntary: Firm initiated
Z-2863-2020Class IITerminatedVoluntary: Firm initiated
Z-2864-2020Class IITerminatedVoluntary: Firm initiated
Z-2865-2020Class IITerminatedVoluntary: Firm initiated
Z-2866-2020Class IITerminatedVoluntary: Firm initiated
Z-2867-2020Class IITerminatedVoluntary: Firm initiated
Z-2868-2020Class IITerminatedVoluntary: Firm initiated
Z-2869-2020Class IITerminatedVoluntary: Firm initiated
Z-2870-2020Class IITerminatedVoluntary: Firm initiated
Z-2871-2020Class IITerminatedVoluntary: Firm initiated