Z-3016-2020Class II

Kit, Pre Op Drah -

Medtronic, Inc. / initiated 2020-07-30 / Terminated / root cause: manufacturing process

Cardinal Health 200, LLC began a recall of KIT, PRE OP DRAH - Product Usage: Preoperative kit used to decolonize intranasal S. aureus on . FDA classified it as Class II and gives the reason as "Only providing one nasal swab in our pre-operative kit instead of the 4 required for proper treatment". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-3016-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
KIT, PRE OP DRAH - Product Usage: Preoperative kit used to decolonize intranasal S. aureus.
Product code
LRP Tray, Surgical
Reason for recall
Only providing one nasal swab in our pre-operative kit instead of the 4 required for proper treatment.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
On July 30th the firm send its consignee a letter notifying them of the field action. Component Part Number: X12048 (SWABSTICK, PVP-I,10%, NASAL DECOLONIZATION, PROFEND,NS) is supplied to Cardinal Health, Presource in a package of 4 nasal swabs. Per the suppliers Instructions for Use, X12048 is a kit that needs to be administered in a 4-swab treatment with two swabs used per nasal passage. Presource is only providing one nasal swab in our pre-operative kit instead of the 4 required for proper treatment. The Presource kit was set up requiring 1 each of the swab component, but confusion in the component set-up caused 1 swab to be kitted instead of 1 package of 4 swabs.
Quantity in commerce
9,150 kits
Distribution
US Nationwide distribution including the state of North Carolina.
Product codes and lots
Model: ORPDH5994A Lot numbers: 558357, 561502,566549, 571467, 574058.  Expiration dates: Varies by kit components.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3016-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3016-2020Class IITerminatedVoluntary: Firm initiated