Z-2958-2020Class II

WallFlexTM Biliary RX Fully Covered Stent System RMV (10mmx60mm Stent) UPN: M00570370 -

Boston Scientific Corporation / initiated 2020-08-07 / Terminated / root cause: packaging

Boston Scientific Corporation began a recall of WallFlexTM Biliary RX Fully Covered Stent System RMV (10mmx60mm Stent) UPN: M00570370 - Product Usage: is indicated for use in the palliative treatment of... on . FDA classified it as Class II and gives the reason as "Mislabeled: WallFlex Biliary RX Uncovered Stent System 10x60mm devices labeled as WallFlexTM Biliary RX Fully Covered Stent System RMV 10x60mm device". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2958-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
WallFlexTM Biliary RX Fully Covered Stent System RMV (10mmx60mm Stent) UPN: M00570370 - Product Usage: is indicated for use in the palliative treatment of biliary strictures produced by malignant neoplasms and relief of malignant biliary obstruction prior to surgery.
Product code
FGE Stents, Drains And Dilators For The Biliary Ducts
Reason for recall
Mislabeled: WallFlex Biliary RX Uncovered Stent System 10x60mm devices labeled as WallFlexTM Biliary RX Fully Covered Stent System RMV 10x60mm device
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Boston Scientific issued Urgent Medical Device Removal - Immediate Action Required on July 31, 2020, via BSC Sales Representatives. The Sales Representatives contacted accounts to determine whether or not the affected inventory had been used. If the product was not used at the site, the product was quarantined. Additionally, customer notification letters to the 32 consignees were sent via Federal Express Priority Overnight mail on July 31st, 2020. Letter states reason for recall, health risk and action to take: Your local Boston Scientific sales representative is available to support this removal. 1. Immediately segregate the affected product to ensure that it will not be used. 2. Contact the Boston Scientific Field Action center for an RGA number to facilitate product return. 3. Ship product back to Boston Scientific We recommend clinical re-evaluation of patients implanted with an uncovered stent from this lot to determine if stent removal is appropriate. 1. Immediately discontinue use of and segregate affected product. 2. Complete the enclosed Reply Verification Tracking Form. 3. Indicate on your Verification Form the quantity of units that you will be returning. 4. Email or fax the Reply Verification Tracking Form to the BSC Field Action Center as described Email: [email protected] or Fax to: Field Action Center 1-866-213-1806. Please email or fax your Reply Verification Tracking Form(s) immediately. Replacement devices will be issued for all product returning to Boston Scientific.
Quantity in commerce
40 units
Distribution
US Nationwide distribution.
Product codes and lots
Lot code:25607019 Exp. Date: June 17th , 2022  GIN:08714729953180

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2958-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2958-2020Class IITerminatedVoluntary: Firm initiated