Z-2899-2020Class II

RayCare, device is stand-alone software -

RaySearch Laboratories AB (publ) / initiated 2020-07-29 / Open, Classified

RAYSEARCH LABORATORIES AB began a recall of RayCare, device is stand-alone software - Product Usage: used to support workflows and scheduling, as well as clinical information, planning and treatment... on . FDA classified it as Class II and gives the reason as "When performing offline image review in RayCare 2C, RayCare 3A and RayCare 3B an offline image review task will guide you to the image viewer with the treatment image related to the fraction and the planning image...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2899-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
RaySearch Laboratories AB (publ)
Product
RayCare, device is stand-alone software - Product Usage: used to support workflows and scheduling, as well as clinical information, planning and treatment management for oncology care.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
When performing offline image review in RayCare 2C, RayCare 3A and RayCare 3B an offline image review task will guide you to the image viewer with the treatment image related to the fraction and the planning image selected. When starting the task, an error message might be displayed stating: Review files not available.
FDA root cause
Under Investigation by firm
Action
Only 1 effected consignee who was emailed a field safety notice on 08/05/2020. This letter instructs the user to: Be aware that registration of a treatment image with a corresponding study with multiple series in the same frame of reference might block the offline image review. Verify the correctness of the data in RayCare PACS before using the data for any decision making. -Please educate staff and all users about this workaround. -Inspect your product and identify all installed units with the above software version number(s). -Confirm you have read and understood this notice by replying to the notification email.
Distribution
US Nationwide distribution including in the state of Tennessee.
Product codes and lots
Model #: 2C SP2, 2C SP3, 3A SP1, 3B and 4A Software Version: 2.2.2.30782 , 2.2.3.53423 , 3.0.1.43597 , 3.1.0.60341 , 4.0.0.60621"

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2899-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2899-2020Class IIOngoingVoluntary: Firm initiated