Z-2971-2020Class II

Cook Medical Universa Soft Ureteral Stent Sets-5 FR 26 cm Reference part Number: UFH-526 -

Cook Incorporated / initiated 2020-07-30 / Terminated / root cause: labeling and UDI

Cook Inc. began a recall of Cook Medical Universa Soft Ureteral Stent Sets-5 FR 26 cm Reference part Number: UFH-526 - Product Usage: Used for temporary internal drainage from the... on . FDA classified it as Class II and gives the reason as "Universa Soft Ureteral Stent Sets were incorrectly placed in the outer boxes labeled as Universa Firm Ureteral Stent Set ;pouch label reflects the packaged product". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2971-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Cook Medical Universa Soft Ureteral Stent Sets-5 FR 26 cm Reference part Number: UFH-526 - Product Usage: Used for temporary internal drainage from the ureteropelvic junction to the bladder
Product code
FAD Stent, Ureteral
Reason for recall
Universa Soft Ureteral Stent Sets were incorrectly placed in the outer boxes labeled as Universa Firm Ureteral Stent Set ;pouch label reflects the packaged product.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
Cook Medical notified William Cook Australia (WCA) on July 30, 2020. WCA notified customers on August 11, 2020. Letter states reason for recall, health risk and action to take: Actions to be Taken by the Customer 1. If affected products have already been placed within a patient, the indwell time should not exceed 180 days. 2. Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all distribution and use of this product. 3. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. Note: Unaffected products that are returned will not be credited. 4. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email ([email protected]). 5. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 6. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to:[email protected].
Quantity in commerce
20 units
Distribution
International distribution to the country of Australia.
Product codes and lots
Lot Number: 13060084

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2971-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2971-2020Class IITerminatedVoluntary: Firm initiated