Z-0060-2021Class II

DJO EMPOWR Tibial Knee Punch Handle Part No. 801-05-040 and 801-05-018 contained in Instrument Tray for...

Encore Medical, LP / initiated 2020-08-03 / Open, Classified

Encore Medical, LP began a recall of DJO EMPOWR Tibial Knee Punch Handle Part No. 801-05-040 and 801-05-018 contained in Instrument Tray for EMPOWR Porous Knee - Product Usage: The Tibial Punch... on . FDA classified it as Class II and gives the reason as "There is a potential for the retaining pin of the tibial punch handle to become loose and fall out while impacting. If undetected, there is a potential of leaving debris in the joint space". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0060-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Encore Medical, LP
Product
DJO EMPOWR Tibial Knee Punch Handle Part No. 801-05-040 and 801-05-018 contained in Instrument Tray for EMPOWR Porous Knee - Product Usage: The Tibial Punch Handle is used in conjunction with one of three appropriately sized Tibial Punch Guides to broach the tibial canal in preparation for the tibial baseplate keel.
Product code
MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
Reason for recall
There is a potential for the retaining pin of the tibial punch handle to become loose and fall out while impacting. If undetected, there is a potential of leaving debris in the joint space.
FDA root cause
Process change control
Action
On August 3, 2020, Field Safety Notices were sent to the firm's distributors by email advising them of this field action, informing them that the firm has received two product complaints where an issue has been experienced with the retaining pin becoming loose and falling out while impacting. Distributors were informed that no adverse events have been reported but that there is a potential of leaving debris in the joint space if this issue were to occur and remain undetected. The letter requests distributors to communicate the following information to implanting surgeons to avoid the potential of leaving debris in the joint space: 1. Sales representative is to inspect instrumentation before every surgery per the Instructions For Use. A small object should be pushed against the end of the pin to confirm it is securely fixed in place. 2. The sales rep should warn of the possibility of the loose pin and steps to take to confirm containment of the loose pin in order to mitigate patient risk. 3. After use of the instrument, it should be verified that the pin remains fixed in place. Distributors were informed that when devices that meet specification become available for the 5 impacted lots, a field exchange will occur as part of a reportable recall. Distributors are requested to complete the acknowledgment form and return it by 8-10-2020 to confirm that they have read and understood best practices for use of the knee punch handle, and have disseminated the information to all employees and contracted Sales Representatives. Customers with questions should contact Teffany Hutto by phone at (512) 834-6255 or by email at [email protected]. On 03/12/2021, recall notices were emailed to additional customers informing them of additional affected models/lots.
Quantity in commerce
190 units
Distribution
U.S. Nationwide distribution including in the states of VA, TN, MN, TX, IL, IN, KY, OH, SC, RI, MD, PA, NY, CA, KS, OK, AR, LA, FL, AL, GA, AZ, NV, WA, MS, ME, NV and Puerto Rico.
Product codes and lots
Model/Catalog no. 801-05-040; :Lot/serial No.s 281905L08, 281905L15, 283808L01, 305197L02, 318871L10 Model/Catalog no. 801-05-018, Lot/serial No.s: 215488L08, 236760L01, 228231L01, 241946L02, 259710L01, 314969L01

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0060-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0060-2021Class IIOngoingVoluntary: Firm initiated