FlexLab (FLX) with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-290-20 and Track to...
Inpeco S.A. began a recall of FlexLab (FLX) with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-290-20 and Track to Rack Module FLX-295-10- Intended Laboratory... on . FDA classified it as Class II and gives the reason as "Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2973-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Inpeco S.p.A.
- Product
- FlexLab (FLX) with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-290-20 and Track to Rack Module FLX-295-10- Intended Laboratory Automation system - Product Usage: The FlexLab 3.6 Automation is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The system consolidates analytical instruments, such as the ARCHITECT c8000 System into a unified workstation that perform a variety of instrument specific assays such as Sodium, Potassium and Chloride. Sodium, Potassium and Chloride measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance.
- Product code
- CEM Electrode, Ion Specific, Potassium
- Reason for recall
- Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Inpeco issued Customer Letter dated August 7, 2020 forwarded to the customers either by Inpeco (for the laboratories where Inpeco is directly the distributor) or by the Inpeco distributors. Letter states reason for recall, health risk and action to take: customers impacted by the Field Action have to complete and return the Recall Letter Receipt Confirmation and Implementation Check form attached to the Customer letter within 15 days to Inpeco or the distributor. The customers impacted by the issue will be contacted by Inpeco or the distributors to arrange a visit to upgrade the impacted modules firmware. Contact reference person: Eva Balzarotti - Regulatory Affairs Manager. E-mail: [email protected] Phone: (+41) 919118 224
- Quantity in commerce
- 2 US and 6 OUS
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AK, FL, IA, LA, MA, MS, MT, NC, OH, TX, VA and the countries of AUSTRALIA, AUSTRIA, BELGIUM, DENMARK, FRANCE, GERMANY, ITALY, JAPAN, NETHERLANDS. NORWAY SAUDI ARABIA SOUTH KOREA SWEDEN UNITED ARAB EMIRATES
- Product codes and lots
- Serial Numbers: US: FLX.0133 FLX.0181 OUS: FLX.0048 FLX.0049 FLX.0121 FLX.0155 FLX.0160 FLX.0200
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2973-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2974-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2975-2020 | Class II | Terminated | Voluntary: Firm initiated |