Z-0124-2021Class II

Sofia SARS Antigen FIA Package Insert The Sofia SARS Antigen FIA is a lateral flow immunofluorescent sandwich...

Ortho-Clinical Diagnostics, Inc. / initiated 2020-09-17 / Terminated / root cause: design

Quidel Corporation began a recall of Sofia SARS Antigen FIA Package Insert The Sofia SARS Antigen FIA is a lateral flow immunofluorescent sandwich assay that is used with the Sofia and Sofia 2... on . FDA classified it as Class II and gives the reason as "Instructions for use in the package insert for SARS Antigen FIA test were updated to remove all references to a specific viral transport medium product as well as the use of other viral transport media. Use of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0124-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
Sofia SARS Antigen FIA Package Insert The Sofia SARS Antigen FIA is a lateral flow immunofluorescent sandwich assay that is used with the Sofia and Sofia 2 instrument intended for the qualitative detection of the nucleocapsid protein antigen from SARS-CoV-2 in nasopharyngeal (NP) and nasal (NS) swab specimens directly from individuals who are suspected of COVID-19 by their healthcare provider within the first five days of the onset of symptoms.
Product code
QKP Coronavirus Antigen Detection Test System.
Reason for recall
Instructions for use in the package insert for SARS Antigen FIA test were updated to remove all references to a specific viral transport medium product as well as the use of other viral transport media. Use of the specific viral transport medium products may result in False-positive results. Directly testing from specimens (NS or NP swab) is recommended.
Root cause tag
design
FDA root cause
Device Design
Action
On 09/17/20, "FIELD CORRECTIVE ACTION CORRECTION" notices dated 09/16/20 were emailed to customers. The firm called customers over the phone to convey the correction information to some customers. Customers were asked to take the following actions: 1) Discontinue the use of affected package inserts. 2) Access the updated instructions for use labeling which can be located by going to quidel.com/docs and selecting the Sofia SARS Antigen FIA Package Insert document. 3) In the event that your facility has used Remel viral transport media products (e.g., M4, M4RT, M6), please contact Quidel Technical Support to discuss your results. 4) Confirm receipt of this notice by returning a signed copy to [email protected] Please feel free to contact Technical Support at [email protected] or by calling 800.874.1517 (in the U.S.) with any questions or concerns that you may have regarding this issue.
Quantity in commerce
10537
Distribution
U.S. Nationwide Distribution: NY, KS, TX, AL, FL, MN, CA, IL, MA, MS, PA, GA, NH, MO, VA, MI, OR, WV, OK, NV, CT, TN, MD, LA, AR, HI, NJ, KY, VT, IA, RI, NM, OH, IN, AK, SC, NC, AZ, SD, NE, WI, and.CO.
Product codes and lots
1438900EN00 (03/20) and 1438901EN00 (05/20)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0124-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0124-2021Class IITerminatedVoluntary: Firm initiated