Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2981-2020Class II

Elecsys Anti-TPO Assay on cobas e analyzers 411, 601, 602, Catalog Number 06368590190

Roche Diagnostics GmbH / initiated 2020-08-20 / Terminated / root cause: manufacturing process

Roche Diagnostics Operations, Inc. began a recall of Elecsys Anti-TPO Assay on cobas e analyzers 411, 601, 602, Catalog Number 06368590190 on . FDA classified it as Class II and gives the reason as "Recent internal verification studies found the current RF interference claim to no longer be met for the assay on the cobas e 411, 601, and 602 analyzers. Therefore, the claim has been updated". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2981-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Elecsys Anti-TPO Assay on cobas e analyzers 411, 601, 602, Catalog Number 06368590190
Product code
JZO System, Test, Thyroid Autoantibody
Reason for recall
Recent internal verification studies found the current RF interference claim to no longer be met for the assay on the cobas e 411, 601, and 602 analyzers. Therefore, the claim has been updated.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Urgent Medical Device Correction notification letters dated 8/20/20 were distributed to customers. Actions Required " Be aware of the new RF interference claims listed in this Urgent Medical Device Correction (UMDC). This UMDC serves as labeling until the updated method sheet (V3), anticipated in September 2020, is available. " Please consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. Customers do not need to perform an additional RF test unless otherwise determined in consultation with the physician or pathologist at your facility. " Complete the attached fax back form (TP-01048) and fax or email it according to the instructions on the form. " File this UMDC for future reference. Please consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. The Elecsys Anti-TPO method sheet is being updated with the new claim for Rheumatoid Factor (RF) interference. Until the updated method sheet (V3) is available, this Urgent Medical Device Correction serves as your labeling. When these method sheets are updated, they will be available on the diagnostics.roche.com website and on the cobas e-library. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
Quantity in commerce
11,870 reagent kits
Distribution
US Nationwide
Product codes and lots
All current and upcoming lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2981-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2981-2020Class IITerminatedVoluntary: Firm initiated