Z-3022-2020Class II

RSP Standard HUMERAL SOCKET INSERT, 40 MM plus 4MM, HXe-Plus, Part No. 509-00-440

Encore Medical, LP / initiated 2020-08-07 / Open, Classified / root cause: labeling and UDI

Encore Medical, LP began a recall of RSP Standard HUMERAL SOCKET INSERT, 40 MM plus 4MM, HXe-Plus, Part No. 509-00-440 on . FDA classified it as Class II and gives the reason as "Product may be mislabeled with the size etched onto products differing from the size information shown on patient sticker labels. Risks associated with this issue may include the Surgeon implanting the incorrect size...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-3022-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Encore Medical, LP
Product
RSP Standard HUMERAL SOCKET INSERT, 40 MM plus 4MM, HXe-Plus, Part No. 509-00-440
Product code
PHX Shoulder Prosthesis, Reverse Configuration
Reason for recall
Product may be mislabeled with the size etched onto products differing from the size information shown on patient sticker labels. Risks associated with this issue may include the Surgeon implanting the incorrect size; significant physical impairment, discomfort and/or disruption of treatment of underlying condition; and the patient experiencing dislocation resulting in revision surgery.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
Letters were sent to the firm's Field/Sales Agents by email beginning on August 7, 2020. The letters included instructions to take the following actions by August 21, 2020: 1. Pass the notice on to all those within your organization or to any organization where the potentially affected product has been transferred, 2.Contact Customer Service at 1-800-456- 8696 to place a replacement order and receive a Return Materials Authorization (RMA) number, and to be provided with a shipping label to return the affected product. 3. Return all affected devices using the RMA number, and 4. Report any adverse events related to this issue to Customer Service. Customers with questions should contact Christine Chesnutt at 512-834-6207 or [email protected].
Quantity in commerce
37 units
Distribution
Distributed in the U.S. to: VA, MN, IN, RI, PA, NY, WA, KS, OK, FL, GA and CA
Product codes and lots
Part 509-00-440, Lot Numbers 387P1042A and 387P1045A, UDI: 00888912144599

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3022-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3021-2020Class IIOngoingVoluntary: Firm initiated
Z-3022-2020Class IIOngoingVoluntary: Firm initiated