ImmuTest Drugs of Abuse Cup (AMT Item Number/ AMT Description): IMCA-5AB-W / "ImmuTest Drugs of Abuse Cup...
Ameditech Inc began a recall of ImmuTest Drugs of Abuse Cup (AMT Item Number/ AMT Description): IMCA-5AB-W / "ImmuTest Drugs of Abuse Cup AMP1000+BZO300+COC300+OPI2000+THC50 With Cr, Ni, pH... on . FDA classified it as Class II and gives the reason as "Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-3036-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ameditech, Inc.
- Product
- ImmuTest Drugs of Abuse Cup (AMT Item Number/ AMT Description): IMCA-5AB-W / "ImmuTest Drugs of Abuse Cup AMP1000+BZO300+COC300+OPI2000+THC50 With Cr, Ni, pH, Bl, S.G.", IMC-8BBCE/ "ImmuTest Drugs of Abuse Cup 8 Drug Detection: (COC300/THC50)+AMP1000+BZO300+BUP10+MET500+MTD300+OPI300 NO ADULTERATION", IMCA-8CM5 /"ImmuTest Drugs of Abuse Cup (OXY100/MDMA500)+(PCP25/AMP1000)+COC150+MET500+OPI2000+THC50 with Cr, Ni, pH, Bl, S.G."
- Product code
- DKC Alcohol Control Materials
- Reason for recall
- Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- A Field Corrective Action Notice letter was sent to affected consignees. The Customer notifications commenced on 19-Aug-2020. The consignees are requested to return all non-expired product in their possession to the recalling firm. Method of Notification, 1st contact: Written field corrective action notification letters will be sent via certified mail with delivery tracking to each consignee along with a pre-paid return label to return any product remaining in stock. 2nd & 3rd contact: Telephone, follow up email, or written mail communication will be used as necessary.
- Quantity in commerce
- 27,375 units
- Distribution
- US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
- Product codes and lots
- ALL Lots Within Expiry
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3029-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3030-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3031-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3032-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3033-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3034-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3035-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3036-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3037-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3038-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3039-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3040-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3041-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3042-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3043-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3044-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3045-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3046-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3047-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3048-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3049-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3050-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3051-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.