Z-0054-2021Class II
VERIGENE Processor SP Model No. 10-0000-07. An in vitro diagnostic device intended for processing and...
Luminex Corporation / initiated 2018-12-10 / Open, Classified / root cause: design
Luminex Corporation began a recall of VERIGENE Processor SP Model No. 10-0000-07. An in vitro diagnostic device intended for processing and identifying nucleic acids on . FDA classified it as Class II and gives the reason as "No results/incorrect results due to failure of the hybridization heater". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0054-2021
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Luminex Corporation
- Product
- VERIGENE Processor SP Model No. 10-0000-07. An in vitro diagnostic device intended for processing and identifying nucleic acids
- Product code
- OCC Respiratory Virus Panel Nucleic Acid Assay System
- Reason for recall
- No results/incorrect results due to failure of the hybridization heater.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- On or about 11/19/2018 there was one complaint from a customer regarding a thermal control failure. In response to the complaint the Field Service Engineer replaced hybridization heater and ran all tests as passed. Subsequently, the customer reported the initial service was ineffective. The VERIGENE SP Processor was then removed from the customer site and returned for analysis. No Customer Notification was communicated for this removal.
- Quantity in commerce
- 1 system
- Distribution
- US distribution in MO. No OUS
- Product codes and lots
- Catalogue Number: 10-0000-07 Serial No. 14230205 UDI Code: 00857573006010
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0054-2021 | Class II | Ongoing | Voluntary: Firm initiated |