Z-0052-2021Class II

Dimension LOCI High Sensitivity Troponin I (TNIH) Flex reagent cartridge-IVD-In vitro diagnostic use in the...

Siemens Healthcare Diagnostics Inc. / initiated 2020-08-19 / Terminated

Siemens Healthcare Diagnostics, Inc. began a recall of Dimension LOCI High Sensitivity Troponin I (TNIH) Flex reagent cartridge-IVD-In vitro diagnostic use in the quantitative measurement of cardiac troponin I in... on . FDA classified it as Class II and gives the reason as "Positive bias across the Analytical Measurement Range of the TNIH assay compared to an unaffected control lot. The average bias observed for patient samples using lots EB0255 and EA0178 of the Dimension EXL Troponin...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0052-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Dimension LOCI High Sensitivity Troponin I (TNIH) Flex reagent cartridge-IVD-In vitro diagnostic use in the quantitative measurement of cardiac troponin I in human plasma Catalog Number: RF627, SMN# 10471068
Product code
MMI Immunoassay Method, Troponin Subunit
Reason for recall
Positive bias across the Analytical Measurement Range of the TNIH assay compared to an unaffected control lot. The average bias observed for patient samples using lots EB0255 and EA0178 of the Dimension EXL Troponin compared with a control lot was +25%. A maximum bias of +34% in patient samples around the 99th percentile observed
FDA root cause
Under Investigation by firm
Action
Siemens issued Urgent Medical Device Correction (UMDC # VC-20-03.B.US) and Urgent Field Safety Notice (UFSN # VC-20-03.B.OUS) issued on August 19, 2020 via FedX to US Dimension customers who received the Dimension TNIH reagent lot to notify them of the issue, the risk to health, and the following actions to be taken by the customer: Review the letter with their Medical Director"If they are using TNIH lot EB0255 DO NOT switch to another TNIH lot during serial measurements for a single patient sample. Discontinue use and discard the TNIH reagent lot EB0255 upon receipt of replacement product. Request your no-charge replacement for lot EB0255. Complete and return the Field Correction Effectiveness Check Form attached to the letter within 30 days. Questions, contact your Siemens Remote Services Center or your local Siemens Technical Support representative.
Quantity in commerce
533 US and 1822 OUS
Distribution
US Nationwide distribution.
Product codes and lots
Lot # EB0255 UDI: 00842768037726EB025520091110471068840

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0052-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0052-2021Class IITerminatedVoluntary: Firm initiated