Dimension LOCI High Sensitivity Troponin I (TNIH) Flex reagent cartridge-IVD-In vitro diagnostic use in the...
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension LOCI High Sensitivity Troponin I (TNIH) Flex reagent cartridge-IVD-In vitro diagnostic use in the quantitative measurement of cardiac troponin I in... on . FDA classified it as Class II and gives the reason as "Positive bias across the Analytical Measurement Range of the TNIH assay compared to an unaffected control lot. The average bias observed for patient samples using lots EB0255 and EA0178 of the Dimension EXL Troponin...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0052-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension LOCI High Sensitivity Troponin I (TNIH) Flex reagent cartridge-IVD-In vitro diagnostic use in the quantitative measurement of cardiac troponin I in human plasma Catalog Number: RF627, SMN# 10471068
- Product code
- MMI Immunoassay Method, Troponin Subunit
- Reason for recall
- Positive bias across the Analytical Measurement Range of the TNIH assay compared to an unaffected control lot. The average bias observed for patient samples using lots EB0255 and EA0178 of the Dimension EXL Troponin compared with a control lot was +25%. A maximum bias of +34% in patient samples around the 99th percentile observed
- FDA root cause
- Under Investigation by firm
- Action
- Siemens issued Urgent Medical Device Correction (UMDC # VC-20-03.B.US) and Urgent Field Safety Notice (UFSN # VC-20-03.B.OUS) issued on August 19, 2020 via FedX to US Dimension customers who received the Dimension TNIH reagent lot to notify them of the issue, the risk to health, and the following actions to be taken by the customer: Review the letter with their Medical Director"If they are using TNIH lot EB0255 DO NOT switch to another TNIH lot during serial measurements for a single patient sample. Discontinue use and discard the TNIH reagent lot EB0255 upon receipt of replacement product. Request your no-charge replacement for lot EB0255. Complete and return the Field Correction Effectiveness Check Form attached to the letter within 30 days. Questions, contact your Siemens Remote Services Center or your local Siemens Technical Support representative.
- Quantity in commerce
- 533 US and 1822 OUS
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Lot # EB0255 UDI: 00842768037726EB025520091110471068840
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0052-2021 | Class II | Terminated | Voluntary: Firm initiated |