Z-3055-2020Class II

Pacific Xtreme (PTCA Balloon Dilation Catheter)

Medtronic, Inc. / initiated 2020-08-26 / Terminated / root cause: labeling and UDI

ev3 Inc. began a recall of Pacific Xtreme (PTCA Balloon Dilation Catheter) on . FDA classified it as Class II and gives the reason as "Labeling discrepancy for the Rated Burst Pressure (RBP) value". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-3055-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Pacific Xtreme (PTCA Balloon Dilation Catheter)
Product code
LIT Catheter, Angioplasty, Peripheral, Transluminal
Reason for recall
Labeling discrepancy for the Rated Burst Pressure (RBP) value.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
On August 26th 2020 the firm send its consignees a letter. Customer Instructions: Medtronic records indicate that your facility has received one or more of the listed Pacific Xtreme PTA Balloon Catheter. As a result, Medtronic requests that you immediately take the following actions: 1.Identify and quarantine all affected and unused Pacific Xtreme PTA Balloon Catheters as listed in the enclosed Customer Notification Detail Report. 2.Return/Exchange all affected and unused Pacific Xtreme PTA Balloon Catheters in your inventory to Medtronic.Contact Medtronic Customer Service at 800-716-6700 to initiate a product return/exchange. Your local Medtronic Representative can assist you as necessary in initiating the return of this product. 3.Complete the enclosed Customer Confirmation Form and email to [email protected]
Quantity in commerce
4286 devices
Distribution
Domestic Distribution: AL, AR, AZ, CA, CO, DC, DE, FL, GA, IA, IL,IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX,VA, WA, WI and WV. International Distribution: Canada and Puerto Rico.
Product codes and lots
Model: PCU060150090 Lot Numbers: 217978890, 218622060.  Model: PCU060150130 Lot Numbers: 217466573, 217603335, 217790972, 217860448, 217876552, 217892598, 217951979, 218153799, 218217163, 218424513, 218482262, 218557813, 218639138, 218712737, 218732096, 218745770, 218815063, 218902560, 218998027, 219036544, 219105117, 219113539, 219236001, 219245043, 219369215, 219544768, 219690958.  Model Number: PCU060200090 Lot Number:217978889  Model Number: PCU060200130 Lot Numbers: 217371273, 217466571, 217617124, 217799993, 217876554, 217914347, 217951978,218037674, 218153802, 218247209, 218321788, 218502816, 218569122, 218655092, 218712742, 218723547, 218732101, 218745773, 218815056, 218846687, 218902556, 218994904, 219018137, 219097771, 219105115, 219113534, 219235998, 219245040, 219333058, 219369216, 219377669, 219444491, 219582240, 219598484, 219690961.  Model Number: PCU060250090 Lot Number: 217603333  Model Number: PCU060250130 Lot Numbers: 217371272, 217451265, 217741799, 217799997, 217867045, 217876555, 218153797, 218217158, 218239743, 218247216, 218424511, 218471575, 218502814, 218569123, 218732098, 218745771, 218815057, 218892319, 218994905, 219018141, 219036626, 219113533, 219235999, 219245041, 219290848, 219358711, 219444490, 219666731.  Model Number: PCU060300090 Lot Numbers: 217790974, 218179770.  Model Number: PCU060300130 Lot Numbers: 217466572, 217558443, 217693877, 217728143, 217741797, 217867826, 217892596, 217914351, 218033558, 218079623, 218170536, 218217164, 218247211, 218321790, 218454476, 218569121, 218655090, 218712741, 218723546, 218752665, 218815061, 218892317, 219018140, 219097767, 219105116, 219217703, 219235996, 219290780, 219369212, 219544769, 219582241, 219598482, 219666726.  Expiration dates are from 03/22/2022 to 04/21/2023

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3055-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3055-2020Class IITerminatedVoluntary: Firm initiated