Z-0098-2021Class II

AquaBplus and AquaBplus B2 Water Purification System. For use with a hemodialysis system to remove organic...

Fresenius Medical Care Renal Therapies Group, LLC / initiated 2020-09-01 / Terminated / root cause: manufacturing process

Fresenius Medical Care Holdings, Inc. began a recall of AquaBplus and AquaBplus B2 Water Purification System. For use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants... on . FDA classified it as Class II and gives the reason as "Incorrectly crimped connectors were installed on the main protective earth conductor on these machines. If the machine is moved or maintenance is performed, these crimp connections could become loose and may adversely...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0098-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
AquaBplus and AquaBplus B2 Water Purification System. For use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from the water used for hemodialysis or related therapies. AquaBplus 1500 - Typ USA - 3 x 208V / 60Hz -Fresenius Part Number: 24-1500-1 AquaBplus B2 2000 / B1 2500 (Bplus DUO 2000) - 3 x 208V / 60Hz- Fresenius Part Number: 24-2000-0
Product code
FIP Subsystem, Water Purification
Reason for recall
Incorrectly crimped connectors were installed on the main protective earth conductor on these machines. If the machine is moved or maintenance is performed, these crimp connections could become loose and may adversely affect the electrical safety of the products, potentially resulting in electrical shock to the Biomed Technician or clinic staff.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Fresenius Medical Care issued letter dated 9/1/20 stating reason for recall, health risk and action to take: A Fresenius Medical Care Renal Therapies Group Water Systems Service Specialist will be contacting you to schedule an appointment to check the connection and replace if necessary. In the meantime, we ask that you do not perform any repairs or maintenance on the machine and wait for your scheduled service date. For questions or concerns or if you need service prior to your scheduled service date, please contact Technical Support Services at 888-553-1109.
Quantity in commerce
15 units
Distribution
US distribution to DE, NC, OK, PA, and TX.
Product codes and lots
9BDS1515 08D51544 OBDS1596 OBPS2864 9BDS1529 OBDS1545 OBPS2799 OBPS2884 9BPS2735 OBDS1551 OBPS2800 OBPS2903 9BPS2767 08D51590 OBPS2810

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0098-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0098-2021Class IITerminatedVoluntary: Firm initiated