Philips North America, LLC began a recall of ProxiDiagnost N90 on . FDA classified it as Class II and gives the reason as "Unexpected increase in peak tube potential (kV) will lead to an increased patient radiation dose". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-3023-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- ProxiDiagnost N90
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Unexpected increase in peak tube potential (kV) will lead to an increased patient radiation dose.
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- Philips sent an Urgent Field Safety Notice dated August 5, 2020 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that Philips plans to correct the defect at no cost to customers by implementing a field correction involving the installation of a software update. A Philips Field Service Engineers will schedule an appointment with customers to install the software update.
- Quantity in commerce
- 71
- Distribution
- United States
- Product codes and lots
- All ProxiDiagnost N90 systems are affected.