Z-0059-2021Class II

1500 Flex Heat and Moisture Exchanger, Catalog Reference 150 -

AB Ulax / initiated 2020-09-04 / Terminated / root cause: manufacturing process

AB ULAX began a recall of 1500 Flex Heat and Moisture Exchanger, Catalog Reference 150 - Product Usage: Heat Moisture Exchangers (HM Es) are designed to enhance humification of... on . FDA classified it as Class II and gives the reason as "Firm has received 6 complaints that the Heat Moisture Exchanger (HME) disconnected from the closed respiratory circuit during use. If the HME separates, the ventilator circuit may not provide adequate ventilation to the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0059-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
AB Ulax
Product
1500 Flex Heat and Moisture Exchanger, Catalog Reference 150 - Product Usage: Heat Moisture Exchangers (HM Es) are designed to enhance humification of ventilator-dependent and non-ventilator dependent patients in critical care, home care and anesthesia.
Product code
BYD Condenser, Heat And Moisture (Artificial Nose)
Reason for recall
Firm has received 6 complaints that the Heat Moisture Exchanger (HME) disconnected from the closed respiratory circuit during use. If the HME separates, the ventilator circuit may not provide adequate ventilation to the patient, potentially resulting in lack of oxygen leading to hypoxia or death.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm distributed customer letters dated September 4, 2020, which informed customers of the product issue. Distributors were asked to inform customers of the recall. End users were asked to do the following: Inventory that is Currently In-Use: 1. Please evaluate any currently ventilated patient who may have such a device (any BALLARD Heat Moisture Exchanger (HME) 150) in their airway circuit, to determine if they belong to any of the impacted product lots. 2. If any of your patient(s) do have a potentially impacted device in their airway circuit, you must replace such device with a commercially available alternative as soon as possible. Due to the potential for brain damage or death if a circuit interruption were not noticed for about 4 minutes, this is a critical instruction. If, despite this instruction and at your own discretion , you elect still not to do so, then you must check carefully to ensure that both ends of the device are seated very firmly within the circuit; consider taping the device in place at both ends to avoid any potential for disconnections; and ensure that all alarms are working, monitored and addressed immediately if indicated. 3. Please complete the attached End-User Product Recall Response Form and return a copy of the form by email to: [email protected] or fax: (888) 229-0320. 4. Once the form is received, a representative will contact you to facilitate return of impacted BALLARD Heat Moisture Exchanger (HME) 150 . 5. Avanos will manage the return of the impacted BALLARD Heat Moisture Exchanger (HME) 150. Please respond within five (5) business days of receipt of this letter. Inventory Already Consumed 6. Please complete the attached End-User Product Recall Response Form (Attachment 1) and select the appropriate box to indicate that you have no inventory of impacted products. 7. Please return this form to Avanos by either email: [email protected] or fax: (888)
Quantity in commerce
1,484,100 units
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Sweden, Taiwan, Hong Kong, and Spain.
Product codes and lots
SAP 109381602, Lots: 2018-09-25 2018-11-20 2018-12-10 2018-12-20 2019-01-03 2019-02-20 2019-03-20 2019-04-15 2019-05-30 2019-06-25 2019-08- 15 2019-09-05 2019- 11-05 2019-11-25  ***  SAP 109381603, Lots 2019-12-20 2020-04-30 2020-05-15 2020-05-20 2020-06-15 2020-06-30

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0059-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0059-2021Class IITerminatedVoluntary: Firm initiated