Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0093-2021Class II

Biomet Comprehensive Reverse Shoulder Glenosphere Mini Baseplate with Taper Adapter-Shoulder Prosthesis item...

Zimmer GmbH / initiated 2020-09-02 / Terminated / root cause: packaging

Biomet, Inc. began a recall of Biomet Comprehensive Reverse Shoulder Glenosphere Mini Baseplate with Taper Adapter-Shoulder Prosthesis item: 010000589 - Product Usage: intended to serve as... on . FDA classified it as Class II and gives the reason as "Product potentially being packaged without a taper adapter or baseplate. The package should contain one (1) baseplate and one (1) taper adapter, product may be packaged with two (2) baseplates and no (0) adapters or...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0093-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Biomet Comprehensive Reverse Shoulder Glenosphere Mini Baseplate with Taper Adapter-Shoulder Prosthesis item: 010000589 - Product Usage: intended to serve as the foundation for the reverse shoulder prosthesis head (glenosphere).
Product code
PHX Shoulder Prosthesis, Reverse Configuration
Reason for recall
Product potentially being packaged without a taper adapter or baseplate. The package should contain one (1) baseplate and one (1) taper adapter, product may be packaged with two (2) baseplates and no (0) adapters or with no (0) baseplates and two (2) adapters
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Zimmer Biomet issued Urgent Medical Device Recall Letter dated 9/02/20 to Distributors via email and Hospital Risk Mangers via courier stating reason for recall, health risk and action to take: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have unconsumed affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form must be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected].
Quantity in commerce
3 units
Distribution
US Nationwide distribution including in the states of CO, MN, OH, WY.
Product codes and lots
Lot Numbers:  488150 -UDI (01) 00880304532465 (17) 300623 (10) 488150; and  564110- UDI (01) 00880304532465 (17) 300609 (10) 564110

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0093-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0093-2021Class IITerminatedVoluntary: Firm initiated