Z-0700-2021Class II

Spinning Spiros, Closed Male Luer, REF: 060-CH2000S; SURPLUG ChemoAccess Closed Male Luer, REF: 066-CH2000S...

Abbott Medical / initiated 2020-08-31 / Open, Classified / root cause: manufacturing process

ICU Medical, Inc. began a recall of Spinning Spiros, Closed Male Luer, REF: 060-CH2000S; SURPLUG ChemoAccess Closed Male Luer, REF: 066-CH2000S; SURPLUG ChemoAccess 5" (13cm) Bag Spike Adaptor... on . FDA classified it as Class II and gives the reason as "Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0700-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Abbott Medical
Product
Spinning Spiros, Closed Male Luer, REF: 060-CH2000S; SURPLUG ChemoAccess Closed Male Luer, REF: 066-CH2000S; SURPLUG ChemoAccess 5" (13cm) Bag Spike Adaptor with Closed Male Luer, REF: 066-CH3034; 35" (89 cm) Appx 3.9 ml, 10 Drop Admin Set w/Integrated Clave Drip Chamber, Clave, Spiros w/Red Cap, Bag Hanger, REF: CH3762; 35" (89 cm) Appx 6.6 ml, 10 Drop Admin Set w/Integrated Clave Drip Chamber, 0.2 Micron Low Protein Binding Filter, Clave, Spiros w/Red Cap, Bag Hanger, REF: CH3763
Product code
FPA Set, Administration, Intravascular
Reason for recall
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On 08/31/2020, Medical Device Recall notices were mailed to customers informing them to take the following actions: 1) Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the attached response form. Return the completed response form to the fax number or e-mail address on the form, even if you do not have the affected product. 3) ICU Medical has some lots of unaffected product available today and is actively increasing the amount of available inventory. In the event specific product is unavailable, consider use of the non-spinning Spiros or the ChemoLock CSTD as alternatives. Please contact customer service for product availability. 4) Return affected product using the return label provided with this letter. Contact Stericycle at 1-888-871-7108 (MF, 8am-5pm ET) if you have not received a return label or require additional labels for returning the affected product. 5) If you have distributed the product further, immediately notify your accounts that received the product identified in the Affected Product / Table 1 sections of this notification and ask them to contact Stericycle to obtain a response form. For further inquiries, please contact: Global Complaint Management, 1-844-654-7780 or [email protected]; Customer Service, 1-866-829-9025, option 8 or, [email protected], (M-F, 8am-6pm CT) On 10/07/2020, the Medical Device Recall notice, with an expanded affected device scope, was sent to customers.
Quantity in commerce
33900
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.
Product codes and lots
Lots: 4605964, 4764534, 4903199, 4901115, 4849120, 4865839, 4895024, 4603209, 4866030

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0700-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0699-2021Class IIOngoingVoluntary: Firm initiated
Z-0700-2021Class IIOngoingVoluntary: Firm initiated
Z-0701-2021Class IIOngoingVoluntary: Firm initiated
Z-0702-2021Class IIOngoingVoluntary: Firm initiated
Z-0703-2021Class IIOngoingVoluntary: Firm initiated
Z-0704-2021Class IIOngoingVoluntary: Firm initiated
Z-0705-2021Class IIOngoingVoluntary: Firm initiated
Z-0706-2021Class IIOngoingVoluntary: Firm initiated
Z-0707-2021Class IIOngoingVoluntary: Firm initiated
Z-0708-2021Class IIOngoingVoluntary: Firm initiated
Z-0709-2021Class IIOngoingVoluntary: Firm initiated
Z-0710-2021Class IIOngoingVoluntary: Firm initiated
Z-0711-2021Class IIOngoingVoluntary: Firm initiated