Vicryl Monofilament Violet 4In (10Cm) Usp 10-0 (M0.2) Single-Armed Cs140-6 -
Ethicon, Inc. began a recall of VICRYL MONOFILAMENT VIOLET 4IN (10CM) USP 10-0 (M0.2) SINGLE-ARMED CS140-6 - Product Usage: Intended for ophthalmic soft tissue approximation/ligation. Product... on . FDA classified it as Class II and gives the reason as "Product in the foil package for VICRYL V960 may not match the foil label and sales unit carton. The correct VICRYL suture is size 10-0, 4 inches (10 cm) in length, has monofilament construction, and a single needle. The...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0116-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ethicon, Inc
- Product
- VICRYL MONOFILAMENT VIOLET 4IN (10CM) USP 10-0 (M0.2) SINGLE-ARMED CS140-6 - Product Usage: Intended for ophthalmic soft tissue approximation/ligation. Product Code: V960G
- Product code
- GAM Suture, Absorbable, Synthetic, Polyglycolic Acid
- Reason for recall
- Product in the foil package for VICRYL V960 may not match the foil label and sales unit carton. The correct VICRYL suture is size 10-0, 4 inches (10 cm) in length, has monofilament construction, and a single needle. The incorrect VICRYL suture is size 6-0, 18 inches (45 cm) in length, has braided construction, and two needles
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Ethicon issued Urgent Medical Device Recall Removal letter on 9/09/20 via email to responsible Johnson & Johnson Affiliates in affected countries who will be contacting all their impacted primary and secondary consignees and/or end customers. The letter states reason for recall, health risk and action to take: 1. Examine your inventory immediately to determine if you have product subject to this recall onhand and quarantine such product(s). 2.Remove the product subject to this recall and communicate the issue to relevant operating roomor materials management personnel, or anyone else in your facility who needs to be informed. 3.If any product subject to this recall has been forwarded to another facility, contact that facility toarrange return. Please consider including a copy of this recall letter when communicating. 4.Complete the Business Reply Form (BRF) (Attachment 3) confirming receipt of this notice. Customers are required to return unused impacted VICRYL Suture, Product Code V960G, LotPJ5352 subject to this recall that are in their inventory immediately. To receive credit reimbursement, customers must return product subject to this recall no later than December 1,2020. Any non-affected product and any product returned after the date specified will not receive replacement.
- Quantity in commerce
- 1,032 eaches (OUS only)
- Distribution
- International distribution including the countries of Australia, Canada, Japan, and New Zealand
- Product codes and lots
- Lot/Serial No.: PJ5352 Expiry Date: 07/31/2024 UDI (GTIN): 10705031051324
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0116-2021 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.