Z-0129-2021Class II
System, Nuclear Magnetic Resonance Imaging -
GE Healthcare / initiated 2020-09-08 / Terminated
GE Healthcare, LLC began a recall of System, Nuclear Magnetic Resonance Imaging - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high... on . FDA classified it as Class II and gives the reason as "Potential electrical arc flash within the Power Gradient & RF (PGR) Cabinet". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0129-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- System, Nuclear Magnetic Resonance Imaging - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- Potential electrical arc flash within the Power Gradient & RF (PGR) Cabinet.
- FDA root cause
- Process change control
- Action
- On 09/08/20 the firm sent a Customer Letter to its consignees with the following instructions: You can continue to use the system. In the event the PGR cabinet needs to be serviced, power down the PGR at the Main Disconnect Panel (MDP) then follow MDP Lockout/Tagout from the current released service methods and documentation. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction
- Quantity in commerce
- 30 Devices total
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of FL, KY, NC, NJ, NM, NY, TX, UT and WA. The countries of Canada, Hungary, India, Korea, China, Russia, Taiwan, Puerto Rico and Ubekistan.
- Product codes and lots
- Model: 5352293, 5726207-5, 5741000-25 and 5741000-29 .
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0127-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0128-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0129-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0130-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0131-2021 | Class II | Terminated | Voluntary: Firm initiated |