Z-0138-2021Class II
TWINFIX 5.0MM AB SUTURE ANCHOR-Absorbable Suture anchors Product Number: 7210711
Smith & Nephew, Inc. / initiated 2020-09-09 / Open, Classified / root cause: manufacturing process
Smith & Nephew, Inc. began a recall of TWINFIX 5.0MM AB SUTURE ANCHOR-Absorbable Suture anchors Product Number: 7210711 on . FDA classified it as Class II and gives the reason as "Product packaging improper or incomplete seal of the pouch surrounding the sterile product may result in a breach of the sterile barrier created by the pouch". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0138-2021
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Smith & Nephew, Inc.
- Product
- TWINFIX 5.0MM AB SUTURE ANCHOR-Absorbable Suture anchors Product Number: 7210711
- Product code
- MAI Fastener, Fixation, Biodegradable, Soft Tissue
- Reason for recall
- Product packaging improper or incomplete seal of the pouch surrounding the sterile product may result in a breach of the sterile barrier created by the pouch
- Root cause tag
- manufacturing process
- FDA root cause
- Equipment maintenance
- Action
- Smith & Nephew issued Urgent Medical Device Recall letter (R-2020-19) FSN on September 9, 2020 via email or overnight mail to customers that received affected products. Letter states reason for recall, health risk, and action to take: Required Actions: 1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. a. If you are a sales rep or distributor, you must notify your customers of the field action and ensure that the required actions are complete. b. Include the customers account number and name in the table below. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below. 4. Complete the remainder of the form, sign and send to [email protected]. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact [email protected].
- Quantity in commerce
- 263,874 (total)
- Distribution
- Nationwide Foreign: AE, AT, AU, BE, BM, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DO, EC, EE, ES, FI, FR, GB, GE, GR, HK, HU, IE, IL, IN, IT, JP,KR, LV,MY, NL, NO, PA,PE, PL, PR, PT, RU, SE, SG, SI, TH, TR, TW, UA, & ZA
- Product codes and lots
- Batch Numbers: 2025421, 2038665,
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0136-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0137-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0138-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0139-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0140-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0141-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0142-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0143-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0144-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0145-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0146-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0147-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0148-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0149-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0150-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0151-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0152-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0153-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0154-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0155-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0156-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0157-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0158-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0159-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0160-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0161-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0162-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0163-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0164-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0165-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0166-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0167-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0168-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0169-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0170-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0171-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0172-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0173-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0174-2021 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0175-2021 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.