MediCult Vitrification Cooling Media, Model Number 12284001F
CooperSurgical, Inc. began a recall of MediCult Vitrification Cooling Media, Model Number 12284001F on . FDA classified it as Class II and gives the reason as "There is a discrepancy between the printing on the Product Tyvek label of the individual component packaging. The Chinese characters on the label indicates that it contains an Equilibration Medium (P/N: 12220001) rather...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0100-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CooperSurgical, Inc.
- Product
- MediCult Vitrification Cooling Media, Model Number 12284001F
- Product code
- MQL Media, Reproductive
- Reason for recall
- There is a discrepancy between the printing on the Product Tyvek label of the individual component packaging. The Chinese characters on the label indicates that it contains an Equilibration Medium (P/N: 12220001) rather than the Vitrification Medium (P/N: 12230001) as indicated on the remainder of the label.
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- URGENT: MEDICAL DEVICE FIELD SAFETY NOTICE letters dated 9/8/20 were sent to customers. Our records indicate that you have purchased the affected Product from CooperSurgical. The affected Vitrification Medium Product is of the correct formulation and functionally acceptable per quality release specifications and will work as intended in accordance with the accompanying Instructions For Use. Please discontinue use of and quarantine the Products, complete the attached Acknowledgement and Receipt Form, and return it to CooperSurgical. Once we receive the Acknowledgement Form, we will arrange for Product replacements and returns of affected Product. The replacement Products will be charged at the same price as the original order and your account will be credited back once the affected Products are returned to CooperSurgical. If you do not have affected Product in inventory, please use the same enclosed Form to indicate that and return it to CooperSurgical so that we may document receipt of this Field Safety Notice. The Chinese Food and Drug Administration has been notified of this Field Safety Notice. We sincerely apologize for any inconvenience caused by this Field Safety Notice. CooperSurgical is committed to high quality, safe and effective products. Please feel free to reach us at +001.203.601.5200 ext. 3300.
- Quantity in commerce
- 531 total boxes
- Distribution
- The products were distributed to China.
- Product codes and lots
- Lot: 20230080
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0100-2021 | Class II | Terminated | Voluntary: Firm initiated |