Synapse PACS Software Version 5.6.1 -
Fujifilm Medical Systems U.S.A., Inc. began a recall of Synapse PACS Software Version 5.6.1 - Product Usage: intended for use as a web based application on an off-the shelf PC which meets or exceeds minimum... on . FDA classified it as Class II and gives the reason as "FUJIFILM has become aware of the possibility that certain CT studies may report a much higher 3D Sphere MAX value than expected. This issue is specific to 3D Sphere - other density tools are not affected". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0282-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FUJIFILM Healthcare Americas Corporation
- Product
- Synapse PACS Software Version 5.6.1 - Product Usage: intended for use as a web based application on an off-the shelf PC which meets or exceeds minimum specifications and is networked with a FUJIFILM Synapse PACS server.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- FUJIFILM has become aware of the possibility that certain CT studies may report a much higher 3D Sphere MAX value than expected. This issue is specific to 3D Sphere - other density tools are not affected.
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- Letter was sent to all affected customers on 09/11/2020 stating the following: FUJIFILM will offer an upgrade to Synapse PACS 7.1.100 to resolve this issue. Synapse PACS 7.1.100 is estimated to be available by the end of September. Until the patch is available, users can continue to safely use Synapse PACS by following these additional instructions: 1. Ensure that the 3D Sphere MAX value is within the spheres range. If the 3D Sphere MAX value is observed to be much higher than expected, the value should not be used when rendering a clinical decision until the issue is corrected. Customers should complete and return the Field Action Verification Form attached to the recall letter.
- Quantity in commerce
- 1
- Distribution
- US Nationwide distribution including in the state of Minnesota.
- Product codes and lots
- Synapse Software Version 5.6.1
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0282-2021 | Class II | Terminated | Voluntary: Firm initiated |