Z-0118-2021Class II
Genesis (GII) -
Smith & Nephew, Inc. / initiated 2020-09-21 / Terminated / root cause: manufacturing process
Smith & Nephew, Inc. began a recall of Genesis (GII) - Product Usage: designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the... on . FDA classified it as Class II and gives the reason as "The anterior locking detail does not meet its design specifications". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0118-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- Genesis (GII) - Product Usage: designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral collateral ligament) are absent or incompetent.
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- The anterior locking detail does not meet its design specifications.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 9/21/20 the firm sent a letter to its consignees with the following instructions: Required Actions: 1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. a. If you are a sales rep or distributor, you must notify your customers of the field action and ensure that the required actions are complete. b. Include the customers account number and name in the table below. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below. 4. Complete the remainder of the form, sign and send to [email protected]. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact [email protected].
- Quantity in commerce
- 47 devices
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of CA, NH, NY, GA, MI, TX, IL, OK, LA, OR, FL, AZ, NJ, MO and the countries of AE. BE, BR, CO, DE, ES, FR, GB, HU, IT, PL, PT, TR.
- Product codes and lots
- Model: 71420966, Lot: 18JY09702
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0118-2021 | Class II | Terminated | Voluntary: Firm initiated |