Z-0298-2021Class II

Siemens Artis one systems with a Display Port transceiver-Fluoroscopic X-Ray System Model: 10848600 -

Siemens Medical Solutions USA, Inc. / initiated 2020-09-21 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Siemens Artis one systems with a Display Port transceiver-Fluoroscopic X-Ray System Model: 10848600 - Product Usage: he examination table as an integrated part... on . FDA classified it as Class II and gives the reason as "Potential malfunction in which the examination room monitor may become intermittently dysfunctional (e.g., no display/flicking/wrong resolution)". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0298-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Artis one systems with a Display Port transceiver-Fluoroscopic X-Ray System Model: 10848600 - Product Usage: he examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions. The Artis one can also support the acquisition of position triggered imaging for spatial data synthesis.
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Potential malfunction in which the examination room monitor may become intermittently dysfunctional (e.g., no display/flicking/wrong resolution).
Root cause tag
design
FDA root cause
Component design/selection
Action
Siemens issued Urgent Medical Device Correction notification letter dated 9/22/20 via AX053/20/S stating reason for recall, health risk and action to take: Siemens will correct the error with a hardware replacement of the Display Port transceiver via Update Instruction AX052/20/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document.
Quantity in commerce
22 units
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, FL, IN, LA, MI, MO, NC, NJ, NY, PA, TX, WA.
Product codes and lots
Serial Numbers: 82010 82038 82039 82040 82044 82054 82065 82066 82070 82102 82105 82108 82109 82205 82213 82217 82267 82308 82354 82355 82378 82383

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0298-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0298-2021Class IITerminatedVoluntary: Firm initiated