Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0313-2021Class II

GE Datex-Ohmeda Aisys CS2 Anesthesia System -

GE Healthcare / initiated 2020-09-17 / Open, Classified / root cause: manufacturing process

GE Healthcare, LLC began a recall of GE Datex-Ohmeda Aisys CS2 Anesthesia System - Product Usage: is intended to provide general inhalation anesthesia and ventilatory support to a wide range of... on . FDA classified it as Class II and gives the reason as "Flow sensors could have damaged tubes in the form of small punctures or cuts". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0313-2021
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Healthcare
Product
GE Datex-Ohmeda Aisys CS2 Anesthesia System - Product Usage: is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric adult). The device is intended for volume or pressure control ventilation.
Product code
BSZ Gas-Machine, Anesthesia
Reason for recall
Flow sensors could have damaged tubes in the form of small punctures or cuts.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
GE will issue an Urgent Medical Device Correction letter to customers with affected flow sensors. This letter will inform the customer of the issue and provide safety instructions, including to inspect all inventory of Flow Sensors. GE Healthcare will replace the Flow Sensors indicated on returned customer response forms, if applicable. In the U.S., this letter will be sent by a traceable means. The letter will be sent to the following titles within the affected accounts: Chief of Anesthesia, Director of Biomedical / Clinical Engineering, and Health Care Administrator / Risk Manager. GE Healthcare will document customer acknowledgment of the receipt and understanding of the letter, and the replacement of Flow Sensors, if needed.
Quantity in commerce
101 devices
Distribution
U.S. states where product was distributed - AL, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI Foreign countries impacted - Australia, Canada, Czech Republic, Estonia, France, Germany, India, Indonesia, Israel, Italy, Kuwait, Latvia, Macedonia, Madagascar, Malaysia, Nicaragua, Panama, Qatar, Russia, Singapore, Slovenia, Spain, Thailand, Ukraine, United Kingdom
Product codes and lots
Devices with installed flow sensors with etched manufacturing date of 2020-08 (August 2020)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0313-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0311-2021Class IIOngoingVoluntary: Firm initiated
Z-0312-2021Class IIOngoingVoluntary: Firm initiated
Z-0313-2021Class IIOngoingVoluntary: Firm initiated
Z-0314-2021Class IIOngoingVoluntary: Firm initiated
Z-0315-2021Class IIOngoingVoluntary: Firm initiated
Z-0316-2021Class IIOngoingVoluntary: Firm initiated
Z-0317-2021Class IIOngoingVoluntary: Firm initiated